GMC-I Messtechnik GmbH
37
Protection Class I Devices
with Terminals for Applied Parts
–
Measurement type PE(TS) - P1
– DUT mains plug to test socket
– Test probe P1 to P1 terminals
Schematic Diagram
Insulation resistance is measured between protective conductor
terminal PE and external, short-circuited applied parts which can
be contacted with test probe P1.
Wiring Diagram
Protection Class I Devices with Outputs for Safety Extra-Low Voltage
–
Measurement type PE(TS) - P1
– DUT mains plug to test socket
– Test probe P1 to P1 terminals
Schematic Diagram
Insulation resistance is measured between the PE terminal and
the safety extra-low voltage outputs, which must be contacted
one after the other with probe P1.
Wiring Diagram
Protection Class I Devices
with Exposed Conductive Parts
–
Measurement type LN(TS) - P1//PE(TS)
– DUT mains plug to test socket
– Test probe P1 to P1 terminals
Schematic Diagram
Insulation resistance is measured between short-circuited mains
terminals (L-N) and external conductive parts which can be con-
tacted with test probe P1 and are
not
connected to the housing,
as well as protective conductor terminal PE at the housing.
Wiring Diagram
Protection Class I Devices
with Terminals for Applied Parts
–
Measurement type LN(TS) - APP
– DUT mains plug to test socket
– APP to APP socket(s)
Schematic Diagram
Wiring Diagram