58542 VXIbus Universal Power Meter
2-56
Publication 21555, Rev. E, September 2002
400
0
0
50
CALIBRATOR DAC
This test steps the calibrator control DAC from 0
to 10 volts in 1 volt steps and measures the
resultant output voltage with the system ADC. It
then compares the results with the limits shown to
the left and if outside any limit it will set the error
number corresponding to that voltage level.
401
1000
950
1050
402
2000
1900
2100
403
3000
2850
3150
404
4000
3800
4200
405
5000
4750
5250
406
6000
5700
6300
407
7000
6650
7350
408
8000
7600
8400
409
9000
8550
9450
410
10000
9500
10500
600
0
50
50
OFFSET AMP 1
This test steps the sensor input amplifier offset
DAC from 0 to 5 volts in 1 volt steps and
measures the resultant output voltage from the
first amplifier stage with the system ADC. It then
compares the results with the limits shown to the
left and if outside any limit it will set the error
number corresponding to that voltage level.
601
1000
950
1050
602
2000
1900
2100
603
3000
2850
3150
604
4000
3800
4200
605
5000
4750
5250
800
0
0
50
GAIN AMP 2
This test steps the sensor input amplifier offset
DAC from 0 volts to 5 volts in 1 volt steps, sets
the second stage gain to 1 and measures the
second stage output voltage with the system
ADC. It then compares the results with the limits
shown to the left and if outside any limit it will set
the error number corresponding to that voltage
level.
801
1000
950
1050
802
2000
1900
2100
803
3000
2850
3150
804
4000
3800
4200
805
5000
4750
5250
1000
0
0
50
THRU FILTER
This test steps the sensor input amplifier offset
DAC from 0 to 5 volts in 1 volt steps, and
measures the filter amplifier output voltage with
the system ADC. It then compares the results with
the limits shown to the left and if outside any limit
it will set the error number corresponding to the
voltage level.
1001
1000
950
1050
1002
2000
1900
2100
1003
3000
2850
3150
1004
4000
3800
4200
1005
5000
4750
5250
Table 2-5: Self-Test Error Indication & Limitations (Continued)
Error Number
Result
Minimum
Maximum
Test Description
Содержание 58542
Страница 3: ...DECLARATION OF CONFORMITY ...
Страница 4: ......
Страница 18: ...58542 VXIbus Universal Power Meter xiv Publication 21555 Rev E September 2002 ...
Страница 20: ...58542 VXIbus Universal Power Meter xvi Publication 21555 Rev E September 2002 ...
Страница 30: ...58452 VXIbus Universal Power Meter 1 10 Publication 21555 Rev E September 2002 ...
Страница 92: ...58542 VXIbus Universal Power Meter 2 62 Publication 21555 Rev E September 2002 ...
Страница 104: ...58452 VXIbus Universal Power Meter 3 12 Publication 21555 Rev E September 2002 ...
Страница 126: ...58452 VXIbus Universal Power Meter 5 8 Publication 21555 Rev E September 2002 ...
Страница 171: ......
Страница 172: ......
Страница 173: ......
Страница 174: ...21360 Rev A has not been u ...
Страница 175: ......
Страница 176: ......
Страница 177: ......
Страница 178: ......
Страница 179: ......
Страница 180: ......
Страница 181: ......
Страница 182: ......
Страница 248: ...58542 VXIbus Universal Power Meter C 2 Publication 21555 Rev E September 2002 ...