58542 VXIbus Universal Power Meter
B-8
Publication 21555, Rev. E, September 2002
B.2.4
Directional Bridges
The 80500 CW Directional Bridges are designed specifically for use with Giga-tronics power meters to
measure the Return Loss/SWR of a test device. Each bridge includes an EEPROM which has been
programmed with Identification Data for that bridge.
The Selection Guide in Table B-5 shows primary specifications. Additional specifications are:
Table B-5: Directional Bridge Selection Guide
Precision CW Return Loss Bridges
Model Freq. Range/
Power Range
Max.
Power
Power Linearity
4
(Frequency >
8
GHz)
Input
Test
Port
Directivity Wgt.
VSWR
80501
10 MHz to 18 GHz
-35 to +20 dBm
+27 dBm
(0.5W)
-35 to +10 dBm
±
0.1 dB
+10 to +20 dBm
±
0.1 dB
±
0.005 dB/dB
Type N(f)
50
Ω
Type N(f)
50
Ω
38 dB
0.340 kg
<1.17:0.01 - 8 GHz
<1.27:8 - 18 GHz
80502
10 MHz to 18 GHz
-35 to +20 dBm
+27 dBm
(0.5 W)
-35 to +10 dBm
±
0.1 dB
+10 to +20 dBm
±
0.1 dB
±
0.005 dB/dB
Type N(f)
50
Ω
APC-7(f)
50
Ω
40 dB
0.340 kg
<1.13:0.01 - 8 GHz
<1.22:8 - 18 GHz
80503
10 MHz to 26.5 GHz
-35 to +20 dBm
+27 dBm
(0.5 W)
-35 to +10 dBm
±
0.1 dB
+10 to +20 dBm
±
0.1 dB
±
0.005 dB/dB
SMA(f)
50
Ω
SMA(f)
50
Ω
35 dB
0.340 kg
<1.22:0.01 - 18 GHz
<1.27:8 - 26.5 GHz
80504
10 MHz to 40 GHz
-35 to +20 dBm
+27 dBm
(0.5 W)
-35 to +10 dBm
±
0.1 dB
+10 to +20 dBm
±
0.1 dB
±
0.005 dB/dB
Type K(f)
50
Ω
Type K(f)
50
Ω
30 dB
0.198 kg
<1.35:0.01 - 26.5 GHz
<1.44:26.5 - 40 GHz
Bridge Frequency Response
Return loss measurements using the 8541/8542 power meter can be frequency compensated using the standard
Open/Short
supplied
with the bridge.
Insertion Loss
6.5 dB, nominal, from input port to test port
Maximum Input Power
+27 dBm (0.5 W)
Directional Bridge Linearity Plus Zero Set & Noise vs. Input Power
(50 MHz, 25
°
C ±5
°
C)
See Table below
Содержание 58542
Страница 3: ...DECLARATION OF CONFORMITY ...
Страница 4: ......
Страница 18: ...58542 VXIbus Universal Power Meter xiv Publication 21555 Rev E September 2002 ...
Страница 20: ...58542 VXIbus Universal Power Meter xvi Publication 21555 Rev E September 2002 ...
Страница 30: ...58452 VXIbus Universal Power Meter 1 10 Publication 21555 Rev E September 2002 ...
Страница 92: ...58542 VXIbus Universal Power Meter 2 62 Publication 21555 Rev E September 2002 ...
Страница 104: ...58452 VXIbus Universal Power Meter 3 12 Publication 21555 Rev E September 2002 ...
Страница 126: ...58452 VXIbus Universal Power Meter 5 8 Publication 21555 Rev E September 2002 ...
Страница 171: ......
Страница 172: ......
Страница 173: ......
Страница 174: ...21360 Rev A has not been u ...
Страница 175: ......
Страница 176: ......
Страница 177: ......
Страница 178: ......
Страница 179: ......
Страница 180: ......
Страница 181: ......
Страница 182: ......
Страница 248: ...58542 VXIbus Universal Power Meter C 2 Publication 21555 Rev E September 2002 ...