58542 VXIbus Universal Power Meter
2-60
Publication 21555, Rev. E, September 2002
Table 2-7: Device Specific Error Messages
Command
Error Message
Example of Problem
ABORt
CALCulate1:DATA?
Burst mode is off
Not burst mode; Command
requires BURSt mode
CALCulate1:DIFFerence 1,1
CALCulate1:RATio 1,1
Conflict in channel configuration
Same sensor
CALCulate1:LIMit:LOWer:UPPer
Conflict between upper and lower limits
Upper is smaller than lower
or lower is greater than
upper
CALCulate1:MAXimum:STATe
Channel is not valid
No sensor or sensor not
calibrated
CALCulate1:MINimum:STATe
Channel is not valid
No sensor or sensor not
calibrated
CALCulate1:MODE
No valid sensor
Channel is not valid
No sensor
No sensor or sensor not
calibrated
CALCulate1:REFerence:COLLect
Normal mode is off
Channel is not valid
Burst mode setup
No sensor or sensor not
calibrated
CALibrate1
No sensor
Sensor not connected to calibrator
Sensor calibration error
No sensor
Calibration error
CALibrate1:ZERO
No sensor
Sensor zeroing error
Zeroing error
DIAGnostic:SENSe1:EEPROM:CORRection
No valid sensor
sensor eeprom data:correction factor number does not
match
No sensor
DIAGnostic:SENSe1:EEPROM:READ
No valid sensor
No sensor
DIAGnostic:SENSe1:EEPROM:CALFSPecial
DIAGnostic:SENSe1:EEPROM:CALFSTandard
No valid sensor
Sensor eeprom data is over space
Sensor eeprom data:cal factor number does not
match
No sensor
Too much data
DIAGnostic:SENSe:EEPROM:FREQSTandard?
No valid sensor
Sensor eeprom data is over space
No sensor
Too much data
DIAGnostic:SENSe1:EEPROM:CALFRange
DIAGnostic:SENSe1:EEPROM:TYPE
No valid sensor
No sensor
DIAGnostic:SENSe1:EEPROM:FREQSPecial
No valid sensor
Sensor eeprom data is over space
Sensor eeprom data:freq is above upper freq limit
Sensor eeprom data:freq is below lower freq limit
Sensor eeprom data:freq number do not match
No sensor
Too much data
DIAGnostic:SENSe1:EEPROM:TYPE
No valid sensor
Sensor eeprom data:upper freq should be larger than
lower freq
No sensor
DIAGnosticSENSe1:EEPROM:WRITe
No valid sensor
Sensor eeprom data:write error
Password is incorrect
No sensor
Sensor eeprom error
Password is activated
Содержание 58542
Страница 3: ...DECLARATION OF CONFORMITY ...
Страница 4: ......
Страница 18: ...58542 VXIbus Universal Power Meter xiv Publication 21555 Rev E September 2002 ...
Страница 20: ...58542 VXIbus Universal Power Meter xvi Publication 21555 Rev E September 2002 ...
Страница 30: ...58452 VXIbus Universal Power Meter 1 10 Publication 21555 Rev E September 2002 ...
Страница 92: ...58542 VXIbus Universal Power Meter 2 62 Publication 21555 Rev E September 2002 ...
Страница 104: ...58452 VXIbus Universal Power Meter 3 12 Publication 21555 Rev E September 2002 ...
Страница 126: ...58452 VXIbus Universal Power Meter 5 8 Publication 21555 Rev E September 2002 ...
Страница 171: ......
Страница 172: ......
Страница 173: ......
Страница 174: ...21360 Rev A has not been u ...
Страница 175: ......
Страница 176: ......
Страница 177: ......
Страница 178: ......
Страница 179: ......
Страница 180: ......
Страница 181: ......
Страница 182: ......
Страница 248: ...58542 VXIbus Universal Power Meter C 2 Publication 21555 Rev E September 2002 ...