Operation
Publication 21555, Rev. E, September 2002
2-45
Syntax:
CALCulate<channel 1 to 2>:LIMit:LOWer
space
<numeric value in dB from -299.99 to 299.99>
Example:
OUTPUT @PWR_MTR;CALC1:LIM:LOW -50.0
! SETS CHANNEL 1 LOWER LIMIT LINE TO -50 DBM
Description:
This command specifies the lower limit line power level. The value should allow for any offset
values currently in use. The default value is -299.99 dB or dBm.
Syntax:
CALCulate<channel 1 to 2>:LIMit:LOWer?
Example:
OUTPUT @PWR_MTR;CALC1:LIM:LOW?
! QUERIES CHANNEL 1 LOWER LIMIT LINE SETTING
Response:
Response is a single number; the lower limit setting.
Description:
This command queries the value of the lower limit line.
Syntax:
CALCulate<channel 1 to 2>:LIMit:STATe
space
<ON | OFF>
Example:
OUTPUT @PWR_MTR;CALC1:LIM:STAT ON
! ENABLES CHANNEL 1 UPPER AND LOWER LIMIT
! CHECKING
Response:
0 = off; 1 = on
Description:
This command activates and deactivates the limit line checking. The default condition is OFF.
Syntax:
CALCulate<channel 1 to 2>:LIMit:STATe?
Example:
OUTPUT @PWR_MTR;CALC1:LIM:STAT?
! QUERIES CHANNEL 1 ON OR OFF STATUS
Response:
0 = off; 1 = on
Description:
This command reports the activated or deactivated status of limit line checking.
Syntax:
CALCulate<channel 1 to 2>:LIMit:UPPer
space
<numeric value in dBm or dB from -299.99 to 299.99>
Example:
OUTPUT @PWR_MTR;CALC1:LIM:UPP 17.0
! SETS CHANNEL 1 UPPER LIMIT LINE TO 17 DBM
Description:
This command specifies the upper limit line power level. The value should allow for any offset
values currently in use. The default value is 299.99 dB or dBm.
Syntax:
CALCulate<channel 1 to 2>:LIMit:UPPer?
Example:
OUTPUT @PWR_MTR;CALC1:LIM:UPP?
! QUERIES CHANNEL 1 UPPER LIMIT LINE SETTING
Response:
Response is a single value in dBm or dB, the upper limit line value.
Description:
This query returns the value of the upper limit line.
Содержание 58542
Страница 3: ...DECLARATION OF CONFORMITY ...
Страница 4: ......
Страница 18: ...58542 VXIbus Universal Power Meter xiv Publication 21555 Rev E September 2002 ...
Страница 20: ...58542 VXIbus Universal Power Meter xvi Publication 21555 Rev E September 2002 ...
Страница 30: ...58452 VXIbus Universal Power Meter 1 10 Publication 21555 Rev E September 2002 ...
Страница 92: ...58542 VXIbus Universal Power Meter 2 62 Publication 21555 Rev E September 2002 ...
Страница 104: ...58452 VXIbus Universal Power Meter 3 12 Publication 21555 Rev E September 2002 ...
Страница 126: ...58452 VXIbus Universal Power Meter 5 8 Publication 21555 Rev E September 2002 ...
Страница 171: ......
Страница 172: ......
Страница 173: ......
Страница 174: ...21360 Rev A has not been u ...
Страница 175: ......
Страница 176: ......
Страница 177: ......
Страница 178: ......
Страница 179: ......
Страница 180: ......
Страница 181: ......
Страница 182: ......
Страница 248: ...58542 VXIbus Universal Power Meter C 2 Publication 21555 Rev E September 2002 ...