Maintenance
Publication 21555, Rev. E, September 2002
5-5
5.4.2
Diagnostic Test Commands
Table 5-1 lists the VXI Diagnostic commands for testing and adjusting the 58542 VXIbus Universal
Power Meter. A typical example is shown after the command syntax. Some commands are described in
the maintenance/calibration section.
It is necessary to disable channels 1 and 2 from taking measurements before using these diagnostic
commands. This is accomplished with the following commands:
CALC1:STAT OFF
CALC2:STAT OFF
Table 5-1: Diagnostic Commands
Command Syntax
Function
DIAGnostic:ADC?
OUTPUT @PWR_MTR;DIAG:ADC?
Query analog to digital converter reading
(Return: 0
-
32767)
DIAGnostic:AVADC?
space
16
OUTPUT @PWR_MTR;DIAG:AVADC? 16
Query analog to digital reading with 16 average
(Average from 1 to 512) (Return: 0
-
32767)
DIAGnostic:AVOLT?
space
16
OUTPUT @PWR_MTR;DIAG:AVOLT? 16
Query analog to digital reading in volts with 16
average (Average from 1 to 512) (Return: 0
-
10)
DIAGnostic:BNC:ANAlog
space
5.5
OUTPUT @PWR_MTR;DIAG:BNC:ANA 5.5
Set analog BNC port output to 5.5 V (-10
-
10)
DIAGnostic:BNC:TRIGger?
OUTPUT @PWR_MTR;DIAG:BNC:TRIG?
Query EXT trigger input high, 5 V or low, 0.00 V,
status (1: Triggered, 0: No trigger)
DIAGnostic:BNC:VPROpf?
OUTPUT @PWR_MTR;DIAG:BNC:VPRO?
Query V
PROP
F port reading (0
-
32767)
DIAGnostic:CALibrator:ATTenuator
space
10
OUTPUT @PWR_MTR;DIAG:CAL:ATT 10
Set calibrator attenuator to 10 dB (0
-
40)
DIAGnostic:CALibrator:ATTenuator?
OUTPUT @PWR_MTR;DIAG:CAL:ATT?
Query calibrator attenuator value
DIAGnostic:CALibrator:EEPROM
space
100
OUTPUT @PWR_MTR;DIAG:CAL:EEPROM 1800123,30,9,23,6,92,0,123456
Set calibrator EEPROM data: cal fac, s/n, min,
hr, day, month, yr, password flag (0:Disable,
1:Enable), password
DIAGnostic:CALibrator:EEPROM?
OUTPUT @PWR_MTR;DIAG:CAL:EEPROM?
Query calibrator eeprom data
DIAGnostic:CALibrator:POWer
space
-23.5
OUTPUT @PWR_MTR;DIAG:CAL:POW-23.5
Set calibrator power to -23.5 dBm
DIAGnostic:CALibrator:PTEMperature?
OUTPUT @PWR_MTR;DIAG:CAL:PTEM?
Query calibrator temperature (in Celsius)
DIAGnostic:CALibrator:SOURce
space
9.5
OUTPUT @PWR_MTR;DIAG:CAL:SOUR 9.5
Set calibrator internal power to 9.5 dBm (Power
output in dBm assuming perfect 10 dB pad)
DIAGnostic:CALibrator:VAMB?
OUTPUT @PWR_MTR;DIAG:CAL:VAMB?
Query calibrator bridge volts (0
-
10)
DIAGnostic:CTABle?
space
<Sensor 1 or 2>
OUTPUT @PWR_MTR;DIAG:CTAB? 1
Query sensor 1 calibration data
Содержание 58542
Страница 3: ...DECLARATION OF CONFORMITY ...
Страница 4: ......
Страница 18: ...58542 VXIbus Universal Power Meter xiv Publication 21555 Rev E September 2002 ...
Страница 20: ...58542 VXIbus Universal Power Meter xvi Publication 21555 Rev E September 2002 ...
Страница 30: ...58452 VXIbus Universal Power Meter 1 10 Publication 21555 Rev E September 2002 ...
Страница 92: ...58542 VXIbus Universal Power Meter 2 62 Publication 21555 Rev E September 2002 ...
Страница 104: ...58452 VXIbus Universal Power Meter 3 12 Publication 21555 Rev E September 2002 ...
Страница 126: ...58452 VXIbus Universal Power Meter 5 8 Publication 21555 Rev E September 2002 ...
Страница 171: ......
Страница 172: ......
Страница 173: ......
Страница 174: ...21360 Rev A has not been u ...
Страница 175: ......
Страница 176: ......
Страница 177: ......
Страница 178: ......
Страница 179: ......
Страница 180: ......
Страница 181: ......
Страница 182: ......
Страница 248: ...58542 VXIbus Universal Power Meter C 2 Publication 21555 Rev E September 2002 ...