Operation
Publication 21555, Rev. E, September 2002
2-49
2.5.27
Saving & Recalling Configurations
■
*RCL
■
*SAV
The 58542 has 21 instrument state memory registers. Registers 1 through 20 are available for store and
recall. Register 0 contains the previous state of the instrument and can be used to toggle between two
different instrument configuration states.
Instrument configuration can be saved to registers 1 through 20.
Any configuration items which are not listed under the *RST or
PRESet conditions are not savable. Make sure all aspects of the
configuration are savable. For example, sensor power sweep cal-
ibration curves can not be saved in the configuration memory
registers. Sensors must be calibrated to the 58542 power meter
each time a new sensor is attached.
2.5.28
Halting Operation
■
ABORt
*RCL
Syntax:
*RCL
space
<memory location number 0 to 20>
Example:
OUTPUT @PWR_MTR;*RCL 19
! RECALL 58542 REGISTER 19
Description:
Recalls instrument configuration. 0 is the PRESet configuration.
*SAV
Syntax:
*SAVe
space
<memory location number 1 to 20>
Example:
OUTPUT @PWR_MTR;*SAV 20
! SAVES AT 58542 REGISTER 20
Description:
Saves current configuration to memory. Saving a configuration to memory position 0 cannot
be done.
ABORt
Example:
OUTPUT @PWR_MTR;ABOR
! HALTS MEASUREMENT & TRIGGERING
Description:
This command stops operation, but it does not interrupt the completion of the current action.
For example, sensor calibration is not interrupted. Burst mode data collection is not
interrupted.
NOTE:
When using the 8035XA Peak Sensor, if a Time-Out occurs due to the sensor
not triggering (i.e., Level too low) then send “Abort” to clear the meter.
CAUTION
☛
Содержание 58542
Страница 3: ...DECLARATION OF CONFORMITY ...
Страница 4: ......
Страница 18: ...58542 VXIbus Universal Power Meter xiv Publication 21555 Rev E September 2002 ...
Страница 20: ...58542 VXIbus Universal Power Meter xvi Publication 21555 Rev E September 2002 ...
Страница 30: ...58452 VXIbus Universal Power Meter 1 10 Publication 21555 Rev E September 2002 ...
Страница 92: ...58542 VXIbus Universal Power Meter 2 62 Publication 21555 Rev E September 2002 ...
Страница 104: ...58452 VXIbus Universal Power Meter 3 12 Publication 21555 Rev E September 2002 ...
Страница 126: ...58452 VXIbus Universal Power Meter 5 8 Publication 21555 Rev E September 2002 ...
Страница 171: ......
Страница 172: ......
Страница 173: ......
Страница 174: ...21360 Rev A has not been u ...
Страница 175: ......
Страница 176: ......
Страница 177: ......
Страница 178: ......
Страница 179: ......
Страница 180: ......
Страница 181: ......
Страница 182: ......
Страница 248: ...58542 VXIbus Universal Power Meter C 2 Publication 21555 Rev E September 2002 ...