Operation
Publication 21555, Rev. E, September 2002
2-29
2.5.17.1
V
PROP
F Configuration
2.5.17.2
Cal Factor Error Control
•
See Error Messages in Section 2.5.34
•
Selected basic SCPI syntax and execution errors apply to these commands
•
Device-specific errors include the following and other -300 level errors
•
A common device-specific error occurs when the frequency sent to the 58542 in the
SENS#:CORR:FREQ
### command is outside the sensor operating frequency range. For example,
sending SENS1:CORR:FREQ 18.4e9 when an 18 GHz (max) 80301A CW Power Sensor is
attached will yield a device-specific error
Figure 2-8: V
PROP
F Configuration
MEM:SLOP
Syntax:
MEMory[:TABLe]:SLOPe
space
<Volts per Hz>
Example:
OUTPUT @PWR_MTR;MEM:SLOP 1E-9
! SETS V
PROP
F SLOPE TO 1 V/GHZ
Description:
This command sets the V/Hz relationship for automated Cal Factor correction.
Syntax:
MEMory[:TABLe]:SLOPe?
Example:
OUTPUT @PWR_MTR;MEM:SLOP?
! QUERIES V
PROP
F SLOPE VALUE
Response:
The voltage to frequency value set that corresponds to source’s V
PROP
F output.
Description:
This command sets the V/Hz slope relationship for V
PROP
F frequency correction.
Содержание 58542
Страница 3: ...DECLARATION OF CONFORMITY ...
Страница 4: ......
Страница 18: ...58542 VXIbus Universal Power Meter xiv Publication 21555 Rev E September 2002 ...
Страница 20: ...58542 VXIbus Universal Power Meter xvi Publication 21555 Rev E September 2002 ...
Страница 30: ...58452 VXIbus Universal Power Meter 1 10 Publication 21555 Rev E September 2002 ...
Страница 92: ...58542 VXIbus Universal Power Meter 2 62 Publication 21555 Rev E September 2002 ...
Страница 104: ...58452 VXIbus Universal Power Meter 3 12 Publication 21555 Rev E September 2002 ...
Страница 126: ...58452 VXIbus Universal Power Meter 5 8 Publication 21555 Rev E September 2002 ...
Страница 171: ......
Страница 172: ......
Страница 173: ......
Страница 174: ...21360 Rev A has not been u ...
Страница 175: ......
Страница 176: ......
Страница 177: ......
Страница 178: ......
Страница 179: ......
Страница 180: ......
Страница 181: ......
Страница 182: ......
Страница 248: ...58542 VXIbus Universal Power Meter C 2 Publication 21555 Rev E September 2002 ...