Operation
Publication 21555, Rev. E, September 2002
2-43
2.5.24
Min/Max Configuration & Monitoring
■
CALCulate:MAXimum
■
CALCulate:MINimum
MIN and MAX channel monitoring records the minimum and maximum variation of channel
amplitude over time. Time zero is set by setting the MIN or MAX STATe to ON. The default Preset and
turn on configuration is ON. After setting to ON, the minimum and maximum value of the measured
value on that channel will be recorded. To reset the MIN and MAX value back to the current value, set
the STATe to ON again. The STAT OFF does not need to be sent during reset of MIN and MAX values.
Individual sensor minimum and maximum power level variations can be monitored only if that sensor is
defined directly to a channel with POW 1 or POW 2. MIN and MAX channel monitoring is only
available in Normal Mode. BURSt and SWIFt modes do not perform MIN and MAX monitoring.
CALC:MAX
Syntax:
CALCulate<channel 1 to 2>:MAXimum:STATe
space
<ON | OFF>
Example:
OUTPUT @PWR_MTR;CALC2:MAX:STAT ON
! ENABLES CHANNEL 2 MAXIMUM TRACKING
Description:
ON activates the MAX monitoring function. The maximum value of the channel (not sensor) is
monitored until the value is reset by CALC#:MAX:STAT ON or CALC#:MAX:STAT OFF is
sent.
Syntax:
CALCulate<channel 1 to 2>:MAXimum:STATe?
Example:
OUTPUT @PWR_MTR;CALC2:MAX:STAT?
! QUERIES CHANNEL 2 MAXIMUM MODE
Response:
1 is for ON or 0 for OFF.
Description:
Reports ON or OFF status of MAX monitoring.
Syntax:
CALCulate<channel 1 to 2>:MAXimum[:MAGnitude]?
Example:
OUTPUT @PWR_MTR;CALC2:MAX?
! QUERIES CHANNEL 2 MAXIMUM VALUE IN DBM
Response:
Highest power reading since CALC2:MAX:STAT ON was sent.
Description:
This command reports the value maintained as the maximum in the 58542 max monitor.
CALC:MIN
Syntax:
CALCulate<channel 1 to 2>:MINimum:STATe
space
ON | OFF
Example:
OUTPUT @PWR_MTR;CALC1:MIN:STAT ON
! ENABLES CHANNEL 1 MINIMUM TRACKING
Description:
This command activates the minimum channel value monitor.
Syntax:
CALCulate<channel 1 to 2>:MINimum [:MAGnitude]?
Example:
OUTPUT @PWR_MTR;CALC1:MIN?
! QUERIES CHANNEL MINIMUM VALUE IN DBM
Response:
Lowest power reading since CALC2:MIN:STAT ON was sent.
Description:
This command reports the value maintained as the minimum in the 58542 min monitor.
Содержание 58542
Страница 3: ...DECLARATION OF CONFORMITY ...
Страница 4: ......
Страница 18: ...58542 VXIbus Universal Power Meter xiv Publication 21555 Rev E September 2002 ...
Страница 20: ...58542 VXIbus Universal Power Meter xvi Publication 21555 Rev E September 2002 ...
Страница 30: ...58452 VXIbus Universal Power Meter 1 10 Publication 21555 Rev E September 2002 ...
Страница 92: ...58542 VXIbus Universal Power Meter 2 62 Publication 21555 Rev E September 2002 ...
Страница 104: ...58452 VXIbus Universal Power Meter 3 12 Publication 21555 Rev E September 2002 ...
Страница 126: ...58452 VXIbus Universal Power Meter 5 8 Publication 21555 Rev E September 2002 ...
Страница 171: ......
Страница 172: ......
Страница 173: ......
Страница 174: ...21360 Rev A has not been u ...
Страница 175: ......
Страница 176: ......
Страница 177: ......
Страница 178: ......
Страница 179: ......
Страница 180: ......
Страница 181: ......
Страница 182: ......
Страница 248: ...58542 VXIbus Universal Power Meter C 2 Publication 21555 Rev E September 2002 ...