58452 VXIbus Universal Power Meter
4-4
Publication 21555, Rev. E, September 2002
where:
P
RF
= calibrator output power reference level
V
COMP
= previously recorded value in Step 9
V
1
= previously recorded value in Step 8
V
0
= previously recorded value in Step 6
R = 200W (assuming HP478A-H75 mount)
Calibration Factor = value for the thermistor mount at 50 MHz (traceable to NIST)
12. Verify that the P
CAL
is within the following limits:
1 mW ±0.019 mW (0.981 to 1.019 mW)
For record purposes, the measured value of P
CAL
can be entered on the Performance Verification Test
Data Sheet at the end of this chapter.
Содержание 58542
Страница 3: ...DECLARATION OF CONFORMITY ...
Страница 4: ......
Страница 18: ...58542 VXIbus Universal Power Meter xiv Publication 21555 Rev E September 2002 ...
Страница 20: ...58542 VXIbus Universal Power Meter xvi Publication 21555 Rev E September 2002 ...
Страница 30: ...58452 VXIbus Universal Power Meter 1 10 Publication 21555 Rev E September 2002 ...
Страница 92: ...58542 VXIbus Universal Power Meter 2 62 Publication 21555 Rev E September 2002 ...
Страница 104: ...58452 VXIbus Universal Power Meter 3 12 Publication 21555 Rev E September 2002 ...
Страница 126: ...58452 VXIbus Universal Power Meter 5 8 Publication 21555 Rev E September 2002 ...
Страница 171: ......
Страница 172: ......
Страница 173: ......
Страница 174: ...21360 Rev A has not been u ...
Страница 175: ......
Страница 176: ......
Страница 177: ......
Страница 178: ......
Страница 179: ......
Страница 180: ......
Страница 181: ......
Страница 182: ......
Страница 248: ...58542 VXIbus Universal Power Meter C 2 Publication 21555 Rev E September 2002 ...