58452 VXIbus Universal Power Meter
1-4
Publication 21555, Rev. E, September 2002
1.2
Specifications
Performance specifications describe the 58542 warranted performance, and apply when using the Series
80300A Power Sensors. Typical performance (shown in italics) is non-warranted.
1.2.1
Range
1.2.2
Accuracy
Frequency Range
10 MHz
-
40 GHz
1
Power Range
-70 dBm
-
+47 dBm (100 pW - 50 Watt)
1
Single Sensor Dynamic Range
CW Sensors: 90 dB
1
Peak Power Sensors: 40 dB, Peak
50 dB, CW
1
Calibrator
Power Sweep calibration signal to dynamically linearize the sensors
Frequency
50 MHz nominal
Settability
The 1 mW (0.0 dBm) level in the Power Sweep Calibrator is factory set to
±
0.7% traceable to the National Institute of Standards and
Technology (NIST). Measure within 15 seconds of setting calibrator to 0.0 dBm
Accuracy
±
1.2% worst case for one year, over temperature range of 5 °C
-
35 °C
Connector
Type N(f) connector, 50
W
VSWR
<1.05 dB (Return Loss >33 dB)
Содержание 58542
Страница 3: ...DECLARATION OF CONFORMITY ...
Страница 4: ......
Страница 18: ...58542 VXIbus Universal Power Meter xiv Publication 21555 Rev E September 2002 ...
Страница 20: ...58542 VXIbus Universal Power Meter xvi Publication 21555 Rev E September 2002 ...
Страница 30: ...58452 VXIbus Universal Power Meter 1 10 Publication 21555 Rev E September 2002 ...
Страница 92: ...58542 VXIbus Universal Power Meter 2 62 Publication 21555 Rev E September 2002 ...
Страница 104: ...58452 VXIbus Universal Power Meter 3 12 Publication 21555 Rev E September 2002 ...
Страница 126: ...58452 VXIbus Universal Power Meter 5 8 Publication 21555 Rev E September 2002 ...
Страница 171: ......
Страница 172: ......
Страница 173: ......
Страница 174: ...21360 Rev A has not been u ...
Страница 175: ......
Страница 176: ......
Страница 177: ......
Страница 178: ......
Страница 179: ......
Страница 180: ......
Страница 181: ......
Страница 182: ......
Страница 248: ...58542 VXIbus Universal Power Meter C 2 Publication 21555 Rev E September 2002 ...