Program Examples
Publication 21555, Rev. E, September 2002
A-15
A.4.4
Reading Power Measurements Example 4
The following program configures the power meter for single channel operation and uses the READ#?
command to return power measurement data in NORMal Mode. Using the READ#? command will not
change power meter configuration items under any circumstances. Note that INIT:CONT ON is illegal
when using READ#?. When continuous trigger arming is desired, use INIT:CONT ON with the
FETCh#? command if desired. Also, use TRIG:SOUR IMM when using the READ#? command.
280
ASSIGN @Pwr_mtr to 70101
290
CLEAR @Pwr_mtr
300
WAIT 1
310
OUTPUT @Pwr_mtr;CALC1:MODE NORM! Can ONLY send Configuration Commands in Normal Mode!!
320
OUTPUT @Pwr_mtr;TRIG:SOUR IMM ! TRIG:SOUR must be IMM when using READ#?
330
OUTPUT @Pwr_mtr;INIT:CONT OFF ! INIT:CONT arming must be OFF when using READ#?
340
!
350
OUTPUT @Pwr_mtr;SENS2:AVER:COUN 1! Sets averaging to manual and one sample per reading.
360
OUTPUT @Pwr_mtr;CALC1:POW 2
! Channel 1 configured to measure sensor 2 power
370
!
380
OUTPUT @Pwr_mtr;SENS2:CORR:FREQ 17.4E9
! Applies Cal Factor for 17.4 GHz to measurement data.
390
!
400
FOR I=1 TO 10
410
OUTPUT @Pwr_mtr;READ2?
! Measures Power at sensor.
420
ENTER @Pwr_mtr;Rdg
430
PRINT Rdg
440
NEXT I
450
!
650
END
Содержание 58542
Страница 3: ...DECLARATION OF CONFORMITY ...
Страница 4: ......
Страница 18: ...58542 VXIbus Universal Power Meter xiv Publication 21555 Rev E September 2002 ...
Страница 20: ...58542 VXIbus Universal Power Meter xvi Publication 21555 Rev E September 2002 ...
Страница 30: ...58452 VXIbus Universal Power Meter 1 10 Publication 21555 Rev E September 2002 ...
Страница 92: ...58542 VXIbus Universal Power Meter 2 62 Publication 21555 Rev E September 2002 ...
Страница 104: ...58452 VXIbus Universal Power Meter 3 12 Publication 21555 Rev E September 2002 ...
Страница 126: ...58452 VXIbus Universal Power Meter 5 8 Publication 21555 Rev E September 2002 ...
Страница 171: ......
Страница 172: ......
Страница 173: ......
Страница 174: ...21360 Rev A has not been u ...
Страница 175: ......
Страница 176: ......
Страница 177: ......
Страница 178: ......
Страница 179: ......
Страница 180: ......
Страница 181: ......
Страница 182: ......
Страница 248: ...58542 VXIbus Universal Power Meter C 2 Publication 21555 Rev E September 2002 ...