58542 VXIbus Universal Power Meter
2-48
Publication 21555, Rev. E, September 2002
OUTP:ANAlog
Syntax:
OUTPut[:BNC]:ANAlog[:STATe]
space
<ON | OFF>
Example:
OUTPUT @PWR_MTR;OUTP:ANA ON
! ENABLES ANALOG OUT FUNCTION
Description:
This command activates and deactivates the analog output. The analog output will not
operate unless this control is set to ON.
Syntax:
OUTPut[:BNC]:ANAlog[:STATe]?
Example:
OUTPUT @PWR_MTR;OUTP:ANA?
! QUERIES ANALOG OUT FUNCTION STATUS
Response:
1 for ON or 0 for OFF.
Description:
This query reports the ON or OFF status of the analog output.
Содержание 58542
Страница 3: ...DECLARATION OF CONFORMITY ...
Страница 4: ......
Страница 18: ...58542 VXIbus Universal Power Meter xiv Publication 21555 Rev E September 2002 ...
Страница 20: ...58542 VXIbus Universal Power Meter xvi Publication 21555 Rev E September 2002 ...
Страница 30: ...58452 VXIbus Universal Power Meter 1 10 Publication 21555 Rev E September 2002 ...
Страница 92: ...58542 VXIbus Universal Power Meter 2 62 Publication 21555 Rev E September 2002 ...
Страница 104: ...58452 VXIbus Universal Power Meter 3 12 Publication 21555 Rev E September 2002 ...
Страница 126: ...58452 VXIbus Universal Power Meter 5 8 Publication 21555 Rev E September 2002 ...
Страница 171: ......
Страница 172: ......
Страница 173: ......
Страница 174: ...21360 Rev A has not been u ...
Страница 175: ......
Страница 176: ......
Страница 177: ......
Страница 178: ......
Страница 179: ......
Страница 180: ......
Страница 181: ......
Страница 182: ......
Страница 248: ...58542 VXIbus Universal Power Meter C 2 Publication 21555 Rev E September 2002 ...