Giga-tronics 2400/2500 Microwave Synthesizer Series
4. Programming Interfaces
Programming Manual, Part Number 34783, Rev A, July 2009
121
Table 20 IEEE 488.2 Common Commands
Command
Name
Description
*TST?
Self-Test
Self-Test Query. It returns '0' if the test succeeds, and '1' if
the test fails. The test sets a predefined group of CW
frequencies and power levels. After each frequency and
power is set, the firmware reads the instrument's LOCK/
LEVEL status. If failing the lock/level, the test is failed. In
order to avoid damage to the device the 2400/2500 is
connected to, maximum attenuation is set if it is available,
or the power level is set to minimum for the duration of the
test. The system will be restored to the pre-test condition
upon completion.
*WAI
Wait-to-Continue
Causes the synthesizer to complete all pending tasks before
executing any additional commands.