Giga-tronics 2400/2500 Microwave Synthesizer Series
4. Programming Interfaces
Programming Manual, Part Number 34783, Rev A, July 2009
119
4.4.2.11
SCPI Commands; Unit Subsystem
Table 18 Unit Subsystem
Command Syntax
Description
UNIT:ANGLe RADians|DEGrees
Sets the default suffix that will be assumed for the
numeric argument for phase adjust programming
commands if no suffix is used
UNIT:ANGLe?
Queries the default suffix that will be assumed for
the numeric argument for phase adjust
programming commands if no suffix is used
UNIT:FREQuency HZ|KZ|KHZ|MZ|MHZ|GZ|GHZ
Sets the default suffix that will be assumed for the
numeric argument of all frequency-related
programming commands if no suffix is used
UNIT:FREQuency?
Queries the default suffix that will be assumed for
the numeric argument of all frequency-related
programming commands if no suffix is used
UNIT:TIME Sec|Msec|USec
Sets the default suffix that will be assumed for the
numeric argument of all power level-related
programming commands if no suffix is used
UNIT:TIME?
Queries the default suffix that will be assumed for
the numeric argument of all frequency-related
programming commands if no suffix is used
4.4.2.12
SCPI Commands; Display Subsystem
Table 19 Display Subsystem
Command Syntax
Description
DISPlay:MENU:STATe n
1: Turns the current menu page to “Remote”; 2, 3:
Turns current page On
DISPlay:TEST
ALL|DIAGonal|HORIZontal|NONE|OFF
Test the screen