4.4.2.6 SCPI Commands; Status Subsystem
Table 13 Status Subsystem SCPI Commands
Command Syntax
Description
STATus:OPERation:CONDition?
Queries the contents of the Operation Condition
register
STATus:OPERation:ENABle n
Sets the contents of the Operation Event Enable
register
STATus:OPERation:ENABle?
Queries the contents of the Operation Event
Enable register
STATus:OPERation: [:EVENt]?
Queries the contents of the Operation Event
register
STATus:OPERation:NTRansition n
Defines which bits in the Operation Condition
register will set the corresponding bit in the
Operation Event register on a one to zero state
change.
STATus:OPERation:NTRansition?
Queries which bits in the Operation Condition
register will set the corresponding bit in the
Operation Event register on a one to zero state
change.
STATus:OPERation: PTRansition n
Defines which bits in the Operation Condition
register will set the corresponding bit in the
Operation Event register on a zero to one state
change.
STATus:OPERation: PTRansition?
Queries which bits in the Operation Condition
register will set the corresponding bit in the
Operation Event register on a zero to one state
change.
STATus:PRESet
Sets several Operation and Questionable registers
to known states
STATus:QUEStionable:CONDition?
Returns the value of the Questionable Status
Condition Register. The value returned is a decimal
value representing the current state of the
register.
STATus:QUEStionable:ENABle <ques>
Sets the Questionable Status Enable Register.
Range of <ques> is 0 - 65535