Giga-tronics 2400/2500 Microwave Synthesizer Series
4. Programming Interfaces
Programming Manual, Part Number 34783, Rev A, July 2009
103
Table 12 Source Subsystem – List Mode SCPI Commands
Command Syntax
Description
[SOURce:]LIST:RANGe:STOP <value>
(HZ|KZ|KHZ|MZ|MHZ|GZ|GHZ)
Set the stop frequency for the range to <value>
[SOURce:]LIST:SEQuence:AUTO?
Query list sequence AUTO mode
[SOURce:]LIST:SYNCout <value>
Generate a pulse at a rear panel BNC output
according to <value>
0-none
1-start
2-end
3-all list points
[SOURce:]LIST:SYNCout?
Query the sync out “mode”
[SOURce:]LIST:SYNCout:DELay <value>
Set Sync out delay in uSec.
Min is 50
Max is 1000000 (1 second)
[SOURce:]LIST:SYNCout:DELay?
Query the sync output delay