Diagnostics and Troubleshooting
10-7
April 2009
Soundspeed
(m/s or f/s)
Displays the measured sound
speed of the fluid.
Check the sound speed
of fluid in
Sound Speeds
and Pipe Size Data.
P#up
Displays signal peaks for the
upstream transducer.
100-
900
<100, >900
P#dn
Displays signal peaks for the
downstream transducer.
100-
900
<100, >900
Reynolds #
Displays the Reynolds Num-
ber.
N/A
N/A
K(RE)
Displays the K Factor, based
on the Reynolds Number.
N/A
N/A
Raw Up
Displays raw upstream signal.
N/A
N/A
Demod Up
Displays demodulated
upstream signal.
N/A
N/A
Skan Corr Up
Displays skan correlation
upstream signal.
N/A
N/A
Raw Down
Displays raw downstream
signal.
N/A
N/A
Demod Down
Displays demodulated
downstream signal.
N/A
N/A
Skan Corr
Down
Displays skan correlation
downstream signal.
N/A
N/A
Xmit Corr Up
Displays raw upstream signal
correlated with transmit
signal.
N/A
N/A
Xmit Corr Dn
Displays raw downstream sig-
nal correlated with transmit
signal.
N/A
N/A
Cross Corr
Displays transmit upstream
signal correlated with transmit
downstream signal.
N/A
N/A
Signal up
Displays the signal strength
for the upstream transducer.
50-80
<50
Signal dn
Displays the signal strength
for the downstream
transducer.
50-80
<50
Qup
Displays the signal quality for
the upstream transducer.
±300 or
higher
between
±100
Table 10-2: Diagnostic Parameters (cont.)
Diagnostic
Parameter
Displays
Good
Bad