Servicing the PT878GC
Testing the Watchdog Timer Circuit
The PT878GC includes a watchdog timer circuit. If a software
error causes the meter to stop responding, this circuit
automatically resets the meter. A properly functioning PT878GC
restarts if you run the Watchdog Test.
To start the Watchdog Test, scroll to the Watchdog Test option in
the Test Menu and press
[ENTER]
. The screen appears similar to
Figure 9-23 below.
Figure 9-23: The Watchdog Test Screen
Press
[F2]
(No) to cancel the test and return to the Menu screen, or
press
[F3]
(Yes) to start the test. The PT878GC should go blank
for a few seconds, and then restart. If it does not follow this
sequence, consult GE Sensing.