April 2009
10-6
Diagnostics and Troubleshooting
Displaying Diagnostic Parameters
As part of its measurement menu, the PT878GC offers a list of
diagnostic parameters to aid in troubleshooting in the event of
flowcell, transducer, or electrical problems. You can select any
diagnostic parameter for display as a measurement as discussed
in Chapter 6 on pages 6-7 and 6-8. Table 10-2 below shows all the
available diagnostic parameters and ranges.
Table 10-2: Diagnostic Parameters
Diagnostic
Parameter
Displays
Good
Bad
Δ
T
Displays Delta T, or the differ-
ence between the upstream
and downstream transit times.
N/A
Continuous
large fluctu-
ations of 1
micro sec-
ond or
more.
Amplitude Up
Displays the value for the
amplitude discriminator of the
upstream transducer.
20-28
<20 or > 28.
Amplitude
Down
Displays the value for the
amplitude discriminator of the
downstream transducer.
20-28
fluctua-
tions
< 20 or > 28
fluctuations
T Up
Displays the upstream
transit time of the
ultrasonic signal in micro
seconds.
N/A
N/A
T Down
Displays the downstream tran-
sit time of the ultrasonic signal
in micro seconds.
N/A
N/A
Gain/Up
Displays upstream gain in dB.
N/A
N/A
Gain/Dn
Displays downstream gain in
dB.
N/A
N/A
CNTup
Displays the AGC DAC count
for upstream gain setting.
N.A.
N.A.
CNTdn
Displays the AGC DAC count
for downstream gain setting.
N.A.
N.A.