Programming Site Data
4-23
April 2009
Entering Mass Flow Parameters (cont.)
2.
If you have enabled Mass Flow, the next prompt asks for the
calculation method. Three options are available: static density,
saturated steam, and steam.
a.
Press
[ENTER]
to open the drop-down menu.
b.
Scroll to the appropriate method for your application.
c.
Press
[ENTER]
to confirm the entry.
•
If you select Saturated Steam:
The prompt asks for the Saturated Steam input. Press
[ENTER]
, scroll to either Temperature or Pressure, and press
[ENTER]
.
•
If you select Static Density:
The prompt asks for the Static Density value. Press
[ENTER]
to open the text box, enter the appropriate value, and press
[ENTER]
to confirm the entry.
Pressing the [
S
] key returns the meter to the Mass Flow prompt,
and then to the Mass Flow tab.
•
To confirm the entries and return to Operate mode, press
[F3]
(OK).
•
To leave the window without confirming the entries and return
to Operate mode, press
[F2]
(Cancel) or the
[ESC]
key.