April 2009
9-46
Servicing the PT878GC
Updating Software Via IrCOMM
Note:
While Windows 95/98/98SE and NT (with QuickBeam)
support IrCOMM, Windows 2000 and XP do not. Refer to
Appendix B.
Before you install new coding, you must be sure that the PC has
the correct protocols to transmit the software to the meter.
Setting up the PC
Note:
The use of
Hyperterminal
is shown here as an example. If
you are using a different communications software, see its
manual for detailed instructions.
1.
From the
Start Menu
,
click
Programs/Accessories/
Hyperterminal
to open the Hyperterminal window
.
2.
If the call is not connected, click on
Connect
from the Call
menu in Hyperterminal.
3.
From the File menu, click
Properties.
Be sure the
New
Connection Properties
window is set to your virtual infrared
port.
Note:
You can determine which virtual port your infrared is
using by opening
Infrared Monitor
in the
Control Panel
and clicking on the
Options
tab. It displays the message,
“Providing Application Support on:” and then lists the
virtual infrared port, as shown in Figure 9-29 on the next
page.