Programming Site Data
4-55
April 2009
Entering a Calibration Factor (cont.)
b.
Enter the desired measurement and units as described on
page 4-40.
c.
Press
[F3]
(OK) to confirm your selection.
4.
The prompt then asks if you wish to edit the K factor table.
a.
Press
[ENTER]
to open the table, which appears similar to
Figure 4-27 below.
Figure 4-27: The KFactor Table
b.
Use the numeric keys to enter the first value for the data
source, and press
[ENTER]
to confirm the entry.
c.
Press the [
X
] key to move to the KFactor column. Use the
numeric keys to enter the corresponding value, and press
[ENTER]
to confirm the entry.
d.
Repeat steps b and c for the remainder of the table.
e.
When you have completed entering values, press
[F3]
(OK)
to confirm the table and return to the Correction Factors
window.
Inputs
Edit KFactor Table
Cancel
OK
0
KFactor
1
2
3
4
5
Data Source