Servicing the PT878GC
9-13
April 2009
Calibrating the Thickness Gauge Transducer (cont.)
a.
Press
[ENTER]
to open the text box.
b.
Use the numeric keys to enter the length.
c.
Press
[ENTER]
to confirm the entry.
4.
Press
[ENTER]
(Calibrate button) to prepare for calibration.
Then apply couplant B to the block surface, and hold the
transducer against the block.
5.
When you have the transducer firmly pressed against the
block, press
[ENTER]
(Ready button) again to perform the
calibration.
6.
Hold the transducer steady until the Set button appears. Press
[ENTER]
to store the calibration.
IMPORTANT:
Be sure to hold the transducer steady until the
GE Sensing wait cursor disappears.
7.
If you have chosen dual calibration, repeat Steps 3 through 6
above for Block 2.
8.
The PT878GC asks for confirmation of the calculated and
current values.
•
Press the [
S
] arrow key to move to the Retry box and press
[ENTER]
to recalculate the value, or
•
Press
[ENTER]
on “Set” to commit the calculated value.
Once you have completed calibration, the transducer should
remain calibrated until the meter memory is cleared or a dramatic
shift in ambient conditions occurs.
To leave the Zero window:
•
Press the [
W
] or [
X
] arrow key to move to another tab to
take readings or perform a velocity calibration.
•
Press
[F2]
(Cancel) to return to Operate Mode without
confirming the sound speed value.
•
Press
[F3]
(OK) to confirm the new value. The PT878GC
returns to Operate Mode.