April 2009
4-32
Programming Site Data
Entering Skan/Correlation Parameters
To open the Skan/Correlation window in the Skan/Measure
option, press the [
X
] arrow key three times (from the All tab) to
reach the Correlate tab, and press
[ENTER]
. The screen appears
similar to Figure 4-14 below.
Figure 4-14: The Correlate Window in the Skan/Measure
Option
To identify the peak of the received signal, the PT878GC uses the
“Threshold” method, in which the peak is identified as the point
where the signal crosses a threshold that is a percentage of the
maximum signal detected.
1.
The first prompt, Min Threshold, asks for the minimum
percentage of peak used to measure transit time.
a.
Press
[ENTER]
to enter the text box.
b.
Use the numeric keys to enter the recommended value.
c.
Press
[ENTER]
to confirm the entry.
Integrate
Skan/Measure
Cancel
OK
All
Min Threshold
Percent Peak
%
Count
Correlate
-100
Measure Only
100
-40
%
%
Max Threshold
Содержание TransPort PT878GC
Страница 1: ...GE Sensing Inspection Technologies TransPort PT878GC Portable Gas Flowmeter User s Manual ...
Страница 29: ...Initial Setup 2 13 April 2009 Keypad cont Figure 2 7 The TransPort PT878GC Keypad ...
Страница 208: ...April 2009 9 4 Servicing the PT878GC Printing Reports cont Figure 9 3 Printout of a Typical Drive Report ...