![GE Moisture Image 1 Series Скачать руководство пользователя страница 85](http://html.mh-extra.com/html/ge/moisture-image-1-series/moisture-image-1-series_programming-manual_128402085.webp)
Advanced Programming
3-21
June 2003
Entering Settings for a
Cyclic Log (cont.)
9. The pointer then moves to the Interval row. Specify the log
interval in minutes and seconds. (The interval can be as short as 5
sec., or up to 99 min. and 59 sec.)
10.At the Time row, enter the interval duration set — the amount of
time in days, hours, minutes and seconds to complete one
measurement cycle. Then proceed to “Selecting the Channel,
Measurement Mode and Units” below.
Note:
If a conflict exists between the Time and the interval time, the
total Time overrules the selected interval time. To set the total
Time with only the record and interval settings, set the Time to
all zeroes for days, hours, minutes and seconds
(000:00:00:00). If dividing the total Time by the number of
records does not result in uniform intervals, the Series 1 uses
the largest interval that will fit within the total Time, and
recalculates the total Time as the largest fitting interval times
the number of records.
Selecting the Channel,
Measurement Mode and
Units
1. The entry for each log channel and mode appears as ___:____ if
blank. Use the arrow keys to move to the entry to change or enter
data. Press [NO] to erase a previous entry.
2. Press numeric keys 1 to 6 to select the desired channel.
3. Press the MODE menu key. A list of measurement modes appears
on the message line (see Table 2-6 on page 2-31).
Note:
Notice the MODE menu key changes to UNIT. This menu key
switches back and forth depending on the parameters you are
selecting.
4. Move the brackets to the desired measurement mode.
5. Press the UNITS key. A list of units appears on the message line.
6. Select the desired units and press [YES].
Repeat steps 1 to 6 until you have selected the desired parameters.
Press the [START] menu key to activate the log. Then press the
[DONE] menu key to return to the Data Logger Menu. If you wish to
check the parameters for a given log, press the [FILE] menu key in
the Data Logger Menu (Figure 3-10 on page 3-17) and then press the
[SETTINGS] menu key.
IMPORTANT:
You cannot change the log parameters once the log
has started!
Содержание Moisture Image 1 Series
Страница 2: ...Moisture Image Series 1 Hygrometer Programming Manual ...
Страница 7: ...Chapter 1 ...
Страница 18: ...Chapter 2 ...
Страница 61: ......
Страница 62: ...Your passcode is 2719 Please remove this page and put it in a safe place for future reference ...
Страница 63: ...Chapter 3 ...
Страница 97: ...Chapter 4 ...
Страница 101: ...Programming with PanaView 4 3 April 2004 Open the Meter Display Window cont Figure 4 1 The Meter Display Window ...
Страница 103: ...Programming with PanaView 4 5 April 2004 Change Meter Programming cont Figure 4 2 Matrix Display After Programming ...