![GE Moisture Image 1 Series Скачать руководство пользователя страница 51](http://html.mh-extra.com/html/ge/moisture-image-1-series/moisture-image-1-series_programming-manual_128402051.webp)
April 2004
2-32
Basic Programming
Setting Up the Matrix
Format
The matrix format consists of six pages that you can set up to display
any combination of channels, measurement modes and measurement
units. For example, if you have a six channel unit, you can use each
page to display various measurements pertaining to one channel or
you can use each page to display one measurement, such as dew
point, for all six channels.
This section describes how to:
•
switch from a line graph to the matrix format
•
set up boxes to display the desired values
•
set the Series 1 to manually or automatically scroll through each
page
Switching from a Line
Graph to the Matrix
Format
At the Main Menu, press the keys as shown in Table 2-7 below.
(Figure 3-18 on page 3-31 for a menu map.) The screen appears
similar to Figure 2-17 below. Press the DONE menu key until Main
Menu appears on the message line.
*This menu key only appears if you are displaying a line graph.
Figure 2-17: Matrix Screen Menu
Table 2-7: Moving from Matrix Screen Menu to Main Menu
Press the following menu keys:
To enter the:
OPTIONS
Option Menu
SCREEN
Graph/Matrix Screen Menu
MATRIX*
Matrix Screen Menu
GRAPHS CYCLE CONTRAST DONE
Matrix Screen Menu 11:35 01/22/2000
pg: 1
1 Dew Point
-18.0 °C
2 Temperature
+35.9 °C
3 H
2
O
+0.535 ppm
v
4 Oxygen
48.3 ppb
v
27.10 PSIG
5 Pressure
X Unassigned
Содержание Moisture Image 1 Series
Страница 2: ...Moisture Image Series 1 Hygrometer Programming Manual ...
Страница 7: ...Chapter 1 ...
Страница 18: ...Chapter 2 ...
Страница 61: ......
Страница 62: ...Your passcode is 2719 Please remove this page and put it in a safe place for future reference ...
Страница 63: ...Chapter 3 ...
Страница 97: ...Chapter 4 ...
Страница 101: ...Programming with PanaView 4 3 April 2004 Open the Meter Display Window cont Figure 4 1 The Meter Display Window ...
Страница 103: ...Programming with PanaView 4 5 April 2004 Change Meter Programming cont Figure 4 2 Matrix Display After Programming ...