Basic Programming
2-15
April 2004
Entering Delta F Oxygen
Cell Calibration Data
Note:
The information in this section applies to Delta F oxygen cells
only. If you are using some other type of oxygen input, refer to
Entering Auxiliary Input Calibration Data on page 2-19.
Your Delta F cell has been factory calibrated using nitrogen as the
reference background gas. If you want to use the oxygen cell with a
different background gas than the cell was calibrated for, refer to the
section Background Gas Correction Factors for the Delta F Oxygen
Cell in Chapter 2 of the Service Manual to determine the correct
oxygen current multiplier.
Use the procedure below to enter the following data:
•
the probe serial number
•
the zero and span range in microamps (
µ
A)
•
the zero and span range in parts per million by volume (ppm
v
)
ppbr, or percent (%)
Note:
Oxygen cells can be ordered to measure in ppm or percent.
Select the proper oxygen cell units in the Probe Configuration
Menu. The units you select will automatically appear in the
second column.
Procedure for Entering Delta F Cell Calibration Data
Referring to the Calibration Data Sheets, enter calibration data
separately for each probe on the designated channel as described
below. Figure 3-19 on page 3-32 for a menu map.
IMPORTANT:
Staple Calibration Data Sheets to the Program
Information List (Appendix A of the Startup Guide)
and store them in a safe place. If they are lost,
contact GE Panametrics for a duplicate. If you alter
any of the calibration data, make sure you make the
change on the Calibration Data Sheet.
1. At the System Calibration Menu (see Table 2-3 on page 2-12),
press the PROBE menu key until the Oxygen Probe Calibration
screen appears. See Figure 2-8.
2. Use the CHANNEL menu key to cycle to the desired channel.
The channel number is indicated in the top right-hand-corner of
the screen. The screen will only display installed channels.
3. Move the pointer to the
S/N
line and press [YES].
Содержание Moisture Image 1 Series
Страница 2: ...Moisture Image Series 1 Hygrometer Programming Manual ...
Страница 7: ...Chapter 1 ...
Страница 18: ...Chapter 2 ...
Страница 61: ......
Страница 62: ...Your passcode is 2719 Please remove this page and put it in a safe place for future reference ...
Страница 63: ...Chapter 3 ...
Страница 97: ...Chapter 4 ...
Страница 101: ...Programming with PanaView 4 3 April 2004 Open the Meter Display Window cont Figure 4 1 The Meter Display Window ...
Страница 103: ...Programming with PanaView 4 5 April 2004 Change Meter Programming cont Figure 4 2 Matrix Display After Programming ...