April 2004
2-22
Basic Programming
Entering Moisture
Reference Values
Use the steps below to change the reference values for moisture
inputs. (Figure 3-19 on page 3-32 for a menu map.)
Caution!
If you modified the supplied cables or are not using GE
Panametrics-supplied cables for moisture probes, you
may be required to adjust the moisture reference data
as described in Performing an MH Calibration/
Adjustment Test on page 1-20 of the Service Manual.
IMPORTANT:
You should record reference data on the Program
Information List supplied in Appendix A of the
Startup Guide, or on a separate sheet of paper, and
keep it in a safe place.
1. At the Reference Menu (see Table 2-4 on page 2-21), press the
PROBE menu key until the Moisture Reference Table appears
(see Figure 2-11 below).
Figure 2-11: Reference Menu for Moisture
2. Compare the data on the screen to the reference data printed on the
label placed on the side or front of the unit, or supplied with a
replacement channel card. If the data is incorrect, use the
procedure below to correct it. If the data is correct, do one of the
following:
•
continue to enter reference values for oxygen cell and pressure
inputs as described in the following sections.
•
press the DONE menu key until Main Menu appears on the
message line; then refer to Displaying Measurements on
page 2-28. When you exit, the Series 1 automatically calibrates
each channel.
HELP INTERVAL PROBE DONE
MOISTURE REFERENCE TABLE
Ch HIGH LOW
1
2
3
4
5
6
AutoCal Interval: (HH.MM) 01:00
Reference Menu
1
Содержание Moisture Image 1 Series
Страница 2: ...Moisture Image Series 1 Hygrometer Programming Manual ...
Страница 7: ...Chapter 1 ...
Страница 18: ...Chapter 2 ...
Страница 61: ......
Страница 62: ...Your passcode is 2719 Please remove this page and put it in a safe place for future reference ...
Страница 63: ...Chapter 3 ...
Страница 97: ...Chapter 4 ...
Страница 101: ...Programming with PanaView 4 3 April 2004 Open the Meter Display Window cont Figure 4 1 The Meter Display Window ...
Страница 103: ...Programming with PanaView 4 5 April 2004 Change Meter Programming cont Figure 4 2 Matrix Display After Programming ...