Chapter 8 • Hydran 201C
i
-1 Installation
8-6
Rev. 6, March 2011
Part 17997
23.
Verify the H201C
i
-1’s display reads 1999 ppm.
Note: The maximum value of the Hydran 201Ci-1’s display scale is 1999 ppm. If the signal
is over scale, the display indicates 1... when the operation mode is set to Force 100 %.
24.
Verify the ammeter reads 20.00 ± 0.40 mA.
25.
Use the
UP
,
DOWN
and
ENTER
keys to force the output to
Off
.
26.
Verify that:
• The H201C
i
-1’s display is blank.
• The analog output drops to 0.00 ± 0.10 mA.
• The System OK indicator (yellow, DS11; item 16 in Figure 3-7 on page 3-12) is off.
27.
Use the
UP
,
DOWN
and
ENTER
keys to return the output to
Normal
.
28.
Verify the signal of the analog output returns to its former value.
29.
When all is done, push
END
.
30.
Disconnect the ammeter.
8.2.4
Verifying the Alarm Indicators and Relays
The steps described in this Section may be used during the Hydran 201C
i
-1 verification,
maintenance and troubleshooting procedures; certain steps may therefore not apply to your
particular situation. Perform all procedures (from Section 8.2.4.1 on page 8-6 to
Section 8.2.4.6 on page 8-9).
The alarm relays are verified by observing the LED’s on the H201C
i
-1’s circuit board and
by connecting an ohmmeter to the main terminal block on the H201C
i
-1 (item 6 in
8.2.4.1
Gas High Alarm
1.
In the Hydran 201T
i
, access the submenu
H201Ti V2.2x;Relays/Analog
.
2.
Access the submenu
Relays/Analog;GasH Relay
.
Содержание Hydran 201i
Страница 1: ...GE Energy Services Digital Energy Hydran Instruction Manual 201i System ...
Страница 6: ...Instruction Manual vi Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Страница 8: ...Instruction Manual viii Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Страница 22: ...Instruction Manual xxii Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Страница 36: ...Chapter 1 Introduction 1 10 Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Страница 72: ...Chapter 3 Hydran 201Ci Controllers 3 16 Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Страница 114: ...Chapter 5 Before Installation 5 16 Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Страница 143: ...Installation Part 17997 Rev 6 March 2011 6 29 Figure 6 9 Installing the CPU Module ...
Страница 176: ...Chapter 8 Hydran 201Ci 1 Installation 8 16 Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Страница 190: ...Chapter 9 Communications and Networking 9 14 Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Страница 216: ...Chapter 11 Operating the Hydran 201i System 11 8 Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Страница 236: ...Chapter 12 Troubleshooting 12 20 Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Страница 244: ...Chapter 13 Maintenance 13 8 Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Страница 254: ...Appendix B Mechanical Drawings B 2 Rev 6 March 2011 Part 17997 Figure B 3 1 5 In Adaptor ...
Страница 255: ...Hydran 201Ti Intelligent Transmitter Part 17997 Rev 6 March 2011 B 3 Figure B 4 1 In Adaptor ...
Страница 262: ...Appendix C Functional Block Diagrams C 4 Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Страница 282: ...Appendix F Default Values of the Parameters F 4 Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Страница 284: ...Appendix G Unit Conversions G 2 Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Страница 298: ...Appendix I Glossary I 12 Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Страница 299: ......