Main and Extended Menus
Part 17997
Rev. 6, March 2011
4-19
•
CheckSum
: Checksum of the above values. This value is added to the list to avoid
erroneous values from being entered.
This submenu does not allow to modify the displayed values. To change the values, use the
NewSensor
submenu (Section 4.5.8 on page 4-25).
The values of the H201 sensor’s ten parameters are stored in the H201T
i
’s nonvolatile
memory.
4.5.4.2
ForceSensor Test
This function is used to test the operation of the Hydran 201 sensor. The Hydran 201T
i
verifies the sensor by injecting a small dc current in the sensor for a few seconds and then
reading the gas level. This test correlates strongly with the sensor response to dissolved
gases and thus allows the H201T
i
to reliably assert the actual condition of the sensor. The
result is immediately interpreted by the microprocessor as follows:
• If the sensor is operating properly, the message
Good
is displayed, but no message is
recorded in the historic data file Events. The H201T
i
’s display returns to the Main
Display mode (see Section 4.2.2 on page 4-4) after the H201 sensor output signal has
returned to normal (two minutes).
• If the sensor test fails (in other words, if an alarm condition is detected), the fail alarm is
triggered, the test result is recorded in the historic data file Events and a message is
displayed on the H201T
i
’s display.
Note: All alarms are ignored for two minutes following the H201 sensor test.
The messages that can be displayed following a sensor test are shown in Table 4-6 on
page 4-20. For the troubleshooting procedure for each of these alarms, see Section 12.1 on
page 12-1.
In addition to the manual tests, a semimonthly, automatic test of the H201 sensor is
performed the first and fifteenth days of each month at midnight. For each automatic test,
certain parameters are recorded in the historic data file Service (see Section 4.5.1.4 on
page 4-14) for further analysis by General Electric Canada.
Note: Each H201Ti is configured for a specific H201 sensor. If you receive more than one
H201Ti’s, take the necessary precautions not to interchange the sensors and H201Ti’s.
Содержание Hydran 201i
Страница 1: ...GE Energy Services Digital Energy Hydran Instruction Manual 201i System ...
Страница 6: ...Instruction Manual vi Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Страница 8: ...Instruction Manual viii Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Страница 22: ...Instruction Manual xxii Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Страница 36: ...Chapter 1 Introduction 1 10 Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Страница 72: ...Chapter 3 Hydran 201Ci Controllers 3 16 Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Страница 114: ...Chapter 5 Before Installation 5 16 Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Страница 143: ...Installation Part 17997 Rev 6 March 2011 6 29 Figure 6 9 Installing the CPU Module ...
Страница 176: ...Chapter 8 Hydran 201Ci 1 Installation 8 16 Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Страница 190: ...Chapter 9 Communications and Networking 9 14 Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Страница 216: ...Chapter 11 Operating the Hydran 201i System 11 8 Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Страница 236: ...Chapter 12 Troubleshooting 12 20 Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Страница 244: ...Chapter 13 Maintenance 13 8 Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Страница 254: ...Appendix B Mechanical Drawings B 2 Rev 6 March 2011 Part 17997 Figure B 3 1 5 In Adaptor ...
Страница 255: ...Hydran 201Ti Intelligent Transmitter Part 17997 Rev 6 March 2011 B 3 Figure B 4 1 In Adaptor ...
Страница 262: ...Appendix C Functional Block Diagrams C 4 Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Страница 282: ...Appendix F Default Values of the Parameters F 4 Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Страница 284: ...Appendix G Unit Conversions G 2 Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Страница 298: ...Appendix I Glossary I 12 Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Страница 299: ......