Chapter 6 • Hydran 201T
i
Installation
6-6
Rev. 6, March 2011
Part 17997
6.2
VERIFICATION OF OPERATION PRIOR TO INSTALLATION
It is strongly recommended to verify the operation of the Hydran 201T
i
’s before instal-
lation in the field (for the installation procedure, see Section 6.3 on page 6-17). The verifi-
cation confirms that no damage occurred during shipping and allows the user to become
familiar with the H201T
i
.
6.2.1
Preparation
1.
Place the Hydran 201T
i
upward on its brass adaptor end.
2.
Unscrew the knob on the front of the aluminum cover.
3.
Remove the cover and place it aside.
4.
Connect a three-wire power cord to the ac Power 3-Terminal Block on top of the I/O
module (item 8 in Figure 2-4 on page 2-8): The live side to
L1/L
, the other side to
L2/N
, and the ground to
E/G
. For additional wiring details, see Table D-1 on
Note: If a supervisory link cable is attached to the H201Ti, do not connect it to the
Hydran 201Ci Controller now. It will be connected later when testing the H201Ci (if
supplied).
5.
Ensure the ac power supply voltage indicated on the label near the terminal block
corresponds to the power source voltage.
6.
Power On the H201T
i
.
CAUTION
WARNING
Strip each wire to a maximum of 8 mm before installing it. Be sure
to insert all strands in the terminal.
CAUTION
WARNING
ELECTRICAL SHOCK HAZARD. Turn off the electric power at the
fuse box or service panel before making any electrical connections,
and ensure a proper ground connection is made before connecting
line voltage. Failure to do so can result in property damage, personal
injury and/or death.
Содержание Hydran 201i
Страница 1: ...GE Energy Services Digital Energy Hydran Instruction Manual 201i System ...
Страница 6: ...Instruction Manual vi Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Страница 8: ...Instruction Manual viii Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Страница 22: ...Instruction Manual xxii Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Страница 36: ...Chapter 1 Introduction 1 10 Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Страница 72: ...Chapter 3 Hydran 201Ci Controllers 3 16 Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Страница 114: ...Chapter 5 Before Installation 5 16 Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Страница 143: ...Installation Part 17997 Rev 6 March 2011 6 29 Figure 6 9 Installing the CPU Module ...
Страница 176: ...Chapter 8 Hydran 201Ci 1 Installation 8 16 Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Страница 190: ...Chapter 9 Communications and Networking 9 14 Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Страница 216: ...Chapter 11 Operating the Hydran 201i System 11 8 Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Страница 236: ...Chapter 12 Troubleshooting 12 20 Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Страница 244: ...Chapter 13 Maintenance 13 8 Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Страница 254: ...Appendix B Mechanical Drawings B 2 Rev 6 March 2011 Part 17997 Figure B 3 1 5 In Adaptor ...
Страница 255: ...Hydran 201Ti Intelligent Transmitter Part 17997 Rev 6 March 2011 B 3 Figure B 4 1 In Adaptor ...
Страница 262: ...Appendix C Functional Block Diagrams C 4 Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Страница 282: ...Appendix F Default Values of the Parameters F 4 Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Страница 284: ...Appendix G Unit Conversions G 2 Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Страница 298: ...Appendix I Glossary I 12 Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Страница 299: ......