Appendix A • Technical Specifications for the Hydran 201
i
System
A-2
Rev. 6, March 2011
Part 17997
Display
Backlit liquid crystal display (LCD); 2 lines x 16 characters
Keypad
6 keys:
ENTER
,
UP
,
DOWN
,
CHANGE
,
ESC
and
END
Alarm Contacts
• Gas High, gas High-High and Fail
• One NO and one NC contacts (type C) per alarm
• 1,250 VA @ 240 Vac; rated current: 5 A
Analog Output
Isolated 4–20 mA; 0–2,000 ppm range; isolation 2,000 V RMS; 10 V load
maximum; 500
@ 20 mA
MISCELLANEOUS
Enclosure
Type NEMA 4X; white, cylindrical aluminum housing; 186 mm (7.3 in)
diameter x 180 mm (7 1/8 in)
Electronic Modules
Totally-encased CPU and I/O modules
Enclosure
Heating/Cooling
325-W heating plate; convection cooling; maintain unit between 15 and
65 °C (59 and 149 °F)
Brass Adaptor
With 1.5-in male NPT threads (standard); mounting on transformer valve;
other dimensions are available
Temperatures
• Oil at the valve: -40 to +90 °C (-40 to +194 °F)
• With finned, high-temperature adaptor (1.5-in male threads only) for oil
temperatures up to 105 °C (221 °F)
• Ambient: -40 to +55 °C (-40 to +131 °F)
• Storage: 5–45 °C (41–113 °F)
Note: At low temperatures, the LCD display might not be legible, but all other
functionalities are fully operational.
Oil Pressure
Vacuum-resistant sensor; withstands 0–700 kPa (0–100 psi)
Power Supply
• Rated voltage: 90–132 Vac or 180–264 Vac, 47–63 Hz
• Rated power: 475 VA maximum
Standards Met
• CISPR 11: Measurement of radiated and conducted emissions
• IEC 60068-2-1: Environmental testing at cold temperature
• IEC 60068-2-2: Environmental testing at dry heat temperature
• IEC 60255-5: Insulation coordination for measuring relays and protection
equipment
• EN 61000-4-2: ESD immunity
• EN 61000-4-3: Radiated electromagnetic field immunity – radio
frequencies
• EN 61000-4-4: Electrical fast transient/burst immunity
• EN 61000-4-5: Surge immunity
• EN 61000-4-6: Immunity to conducted disturbance induced by radio-
frequency fields
• EN 61000-4-8: Power frequency magnetic field immunity
• EN 61000-4-11: Voltage dips, short interruptions and voltage variations
immunity (applicable for ac power supply only)
• EN 61010-1: Product safety
Safety Conditions
Designed to be safe at least under the following conditions:
• Altitude up to 2,000 m
• Enclosure at temperatures ranging from -40 to +55 °C (-40 to +131 °F)
and at relative humidities of 95 % or less
Содержание Hydran 201i
Страница 1: ...GE Energy Services Digital Energy Hydran Instruction Manual 201i System ...
Страница 6: ...Instruction Manual vi Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Страница 8: ...Instruction Manual viii Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Страница 22: ...Instruction Manual xxii Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Страница 36: ...Chapter 1 Introduction 1 10 Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Страница 72: ...Chapter 3 Hydran 201Ci Controllers 3 16 Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Страница 114: ...Chapter 5 Before Installation 5 16 Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Страница 143: ...Installation Part 17997 Rev 6 March 2011 6 29 Figure 6 9 Installing the CPU Module ...
Страница 176: ...Chapter 8 Hydran 201Ci 1 Installation 8 16 Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Страница 190: ...Chapter 9 Communications and Networking 9 14 Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Страница 216: ...Chapter 11 Operating the Hydran 201i System 11 8 Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Страница 236: ...Chapter 12 Troubleshooting 12 20 Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Страница 244: ...Chapter 13 Maintenance 13 8 Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Страница 254: ...Appendix B Mechanical Drawings B 2 Rev 6 March 2011 Part 17997 Figure B 3 1 5 In Adaptor ...
Страница 255: ...Hydran 201Ti Intelligent Transmitter Part 17997 Rev 6 March 2011 B 3 Figure B 4 1 In Adaptor ...
Страница 262: ...Appendix C Functional Block Diagrams C 4 Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Страница 282: ...Appendix F Default Values of the Parameters F 4 Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Страница 284: ...Appendix G Unit Conversions G 2 Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Страница 298: ...Appendix I Glossary I 12 Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Страница 299: ......