Verification of Operation Prior to Installation
Part 17997
Rev. 6, March 2011
6-7
6.2.2
Initial Observations
1.
If the battery was just reconnected, the H201T
i
performs the first power-up sequence
(see Section 4.2.5.3 on page 4-7). Enter the requested parameters (date, time, etc.)
using the H201T
i
’s keypad.
2.
Observe the display and ensure the H201T
i
goes through the self-test sequence shown
Note: The self-test sequence is executed only if the power was down for more than five
minutes. Otherwise, only the message System;Initialization appears.
Note: A sensor test is executed at the end of the self-test sequence only if the scheduled last
sensor test was not performed. In this case, the H201Ti displays the message Sensor Good
and waits 120 seconds before resuming normal display.
3.
Verify that the H201T
i
displays the Main Display mode (the parameters shown in
Table 4-1 on page 4-5, at the rate of one every five seconds).
4.
Verify the operation of the H201T
i
by observing the six small LED diagnostic
indicators on the right side of the I/O module (from left to right in Figure 2-7 on
page 2-12):
•
-
Heating state LED
: This yellow LED flashes once every five seconds in proportion
to the amount of power supplied to the heating plate.
– The LED must remain lit if the internal temperature is below the set point (adjustable
Temp SetPoint
parameter in the
Temperature;DynOil Sampl
submenu).
– The LED must flash if the internal temperature is close to the set point.
– The LED must remain off if the internal temperature exceeds the set point.
•
-5 - Regulated -5 Vdc supply LED
: This dim, green LED must be lit to confirm the
presence of the -5 Vdc power supply.
•
+5 - Reg5 Vdc supply LED
: This dim, green LED must be lit to confirm the
presence of the +5 Vdc power supply.
•
OK - System state indicator LED
: This green LED must be lit to confirm that the
H201T
i
is operating properly and that the fail alarm relay is energized. It turns off if a
fail alarm condition is detected and if the fail alarm relay is de-energized.
•
HH - High-High alarm state LED
: This red LED is lit when a High-High alarm
condition is detected and the High-High alarm relay is energized. This LED must
therefore be off.
C
Содержание Hydran 201i
Страница 1: ...GE Energy Services Digital Energy Hydran Instruction Manual 201i System ...
Страница 6: ...Instruction Manual vi Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Страница 8: ...Instruction Manual viii Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Страница 22: ...Instruction Manual xxii Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Страница 36: ...Chapter 1 Introduction 1 10 Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Страница 72: ...Chapter 3 Hydran 201Ci Controllers 3 16 Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Страница 114: ...Chapter 5 Before Installation 5 16 Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Страница 143: ...Installation Part 17997 Rev 6 March 2011 6 29 Figure 6 9 Installing the CPU Module ...
Страница 176: ...Chapter 8 Hydran 201Ci 1 Installation 8 16 Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Страница 190: ...Chapter 9 Communications and Networking 9 14 Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Страница 216: ...Chapter 11 Operating the Hydran 201i System 11 8 Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Страница 236: ...Chapter 12 Troubleshooting 12 20 Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Страница 244: ...Chapter 13 Maintenance 13 8 Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Страница 254: ...Appendix B Mechanical Drawings B 2 Rev 6 March 2011 Part 17997 Figure B 3 1 5 In Adaptor ...
Страница 255: ...Hydran 201Ti Intelligent Transmitter Part 17997 Rev 6 March 2011 B 3 Figure B 4 1 In Adaptor ...
Страница 262: ...Appendix C Functional Block Diagrams C 4 Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Страница 282: ...Appendix F Default Values of the Parameters F 4 Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Страница 284: ...Appendix G Unit Conversions G 2 Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Страница 298: ...Appendix I Glossary I 12 Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Страница 299: ......