K415 Issue No. 1
52
(5)
Set the input current to 55 ± 0.1mA and enter the measured full-scale input
current:
(6)
Press the
ENTER
key to accept the calibration. Press the
EXIT
key four times to
quit the
CALIBRATION
and
SETUP
modes.
(7)
Verify the instrument calibration by applying the test currents given in Table 3 to
the milliammeter.
(8)
Disconnect the calibration/test equipment.
Table 3 Electrical Current Input Calibration Tolerances
Current Calibration, 55mA Range Verification Tolerances
Applied Current (mA)
Permissible Deviation
-55
± 0.05% rdg ± 0.004% FS
-45
± 0.05% rdg ± 0.004% FS
-35
± 0.05% rdg ± 0.004% FS
-25
± 0.05% rdg ± 0.004% FS
-15
± 0.05% rdg ± 0.004% FS
-5
± 0.05% rdg ± 0.004% FS
0
± 0.05% rdg ± 0.004% FS
5
± 0.05% rdg ± 0.004% FS
15
± 0.05% rdg ± 0.004% FS
25
± 0.05% rdg ± 0.004% FS
35
± 0.05% rdg ± 0.004% FS
45
± 0.05% rdg ± 0.004% FS
55
± 0.05% rdg ± 0.004% FS
Calibration
Содержание Druck DPI 610
Страница 1: ...Druck DPI 610 615 Portable Pressure Calibrator Series User manual K415 GE Sensing ...
Страница 2: ......
Страница 10: ...K415 Issue No 1 viii ...
Страница 69: ...59 K415 Issue No 1 Hydraulic Calibrator Versions ...
Страница 70: ...K415 Issue No 1 60 This page intentionally left blank ...
Страница 76: ...K415 Issue No 1 66 This page intentionally left blank ...
Страница 77: ...67 K415 Issue No 1 Low Pressure Calibrator Versions ...
Страница 78: ...K415 Issue No 1 68 This page intentionally left blank ...
Страница 81: ...71 K415 Issue No 1 C Closed O Open Figure B2 Test Connections Low Pressure Calibrator Versions ...
Страница 87: ...77 K415 Issue No 1 Appendix 1 DATA LOG FILE EXAMPLE ...
Страница 88: ...K415 Issue No 1 78 This page intentionally blank ...