K415 Issue No. 1
62
Safety Instructions
WARNING
HYDRAULIC FLUID IS INJURIOUS. OBSERVE RELEVANT HEALTH AND SAFETY
PRECAUTIONS. USE APPROPRIATE PROTECTIVE BARRIERS AND EYE
PROTECTION.
BEFORE APPLYING PRESSURE, EXAMINE ALL FITTINGS AND EQUIPMENT FOR
DAMAGE AND ENSURE THAT ALL EQUIPMENT IS TO THE CORRECT PRESSURE
RATING.
DO NOT EXCEED THE MAXIMUM WORKING PRESSURE OF THE INSTRUMENT
(INDICATED ON START-UP SCREEN AT SWITCH-ON).
Caution:
Observe absolute cleanliness when using the instrument.
Severe damage can be caused if equipment connected to this
instrument is contaminated. Connect only clean equipment to
the instrument. To avoid any contamination, an external filter
is recommended.
Preparation for Use (Figure A1)
• Use an appropriate seal and connect the bleed hose assembly (3) to the priming port.
• Make sure that the screw-press (5) is wound fully in (clockwise).
• Make sure that the isolation valve (1) is open (fully counter clockwise).
• Use a bonded seal (6) and connect the unit or system under test (7) to the pressure
port.
WARNING
MAKE SURE THAT THE CONNECTING LINE TO THE EXTERNAL DEVICE OR
SYSTEM IS CAPABLE OF WITHSTANDING THE LINE PRESSURE TO BE APPLIED.
Note:
A bleed point must be provided on the external device.
Hydraulic Calibrator Versions Operation
Содержание Druck DPI 610
Страница 1: ...Druck DPI 610 615 Portable Pressure Calibrator Series User manual K415 GE Sensing ...
Страница 2: ......
Страница 10: ...K415 Issue No 1 viii ...
Страница 69: ...59 K415 Issue No 1 Hydraulic Calibrator Versions ...
Страница 70: ...K415 Issue No 1 60 This page intentionally left blank ...
Страница 76: ...K415 Issue No 1 66 This page intentionally left blank ...
Страница 77: ...67 K415 Issue No 1 Low Pressure Calibrator Versions ...
Страница 78: ...K415 Issue No 1 68 This page intentionally left blank ...
Страница 81: ...71 K415 Issue No 1 C Closed O Open Figure B2 Test Connections Low Pressure Calibrator Versions ...
Страница 87: ...77 K415 Issue No 1 Appendix 1 DATA LOG FILE EXAMPLE ...
Страница 88: ...K415 Issue No 1 78 This page intentionally blank ...