![GE Druck DPI 610 Скачать руководство пользователя страница 71](http://html.mh-extra.com/html/ge/druck-dpi-610/druck-dpi-610_user-manual_132657071.webp)
61
K415 Issue No. 1
Introduction (Figure A1)
These versions of the DPI 610 and DPI 615 Calibrators provide manual generation of
hydraulic pressure and consist of a screw-press with a priming pump and isolation valve.
1
Isolation valve 2
Priming port
3
Bleed hose assembly
4
Priming pump
5
Screw-press
6
Bonded seal
7
UUT/system under test
Figure A1 - DPI 610/615 HC Connections
The internal hydraulic parts are brass, stainless steel, copper, nylon and fluorocarbon
rubber (Viton). The hydraulic fluid can be either demineralized water or one of the
hydraulic fluids listed below:
Caution:
Only use the recommended hydraulic fluids.
DO NOT mix hydraulic fluids.
Table A1 - Recommended Hydraulic Fluids
ISO 3448
Viscosity
grade
Approx. SAE
Viscosity
Classification
Shell
Esso
(Exxon)
Mobil
VG10
5W
Tellus R10
Nuto H10
Velocite No. 6
VG15
Tellus T15
Tellus V15
Nuto H15
VG22
Tellus 22
Tellus R22
Nuto H22
DTE 22
VG32
10W
Tellus V32
Nuto H32
DTE Oil Light
DTE 24
VG37
Tellus 37
Tellus R37
Tellus T37
Tellus V37
7
6
5
4
1
2
3
5
Hydraulic Calibrator Versions Operation
Содержание Druck DPI 610
Страница 1: ...Druck DPI 610 615 Portable Pressure Calibrator Series User manual K415 GE Sensing ...
Страница 2: ......
Страница 10: ...K415 Issue No 1 viii ...
Страница 69: ...59 K415 Issue No 1 Hydraulic Calibrator Versions ...
Страница 70: ...K415 Issue No 1 60 This page intentionally left blank ...
Страница 76: ...K415 Issue No 1 66 This page intentionally left blank ...
Страница 77: ...67 K415 Issue No 1 Low Pressure Calibrator Versions ...
Страница 78: ...K415 Issue No 1 68 This page intentionally left blank ...
Страница 81: ...71 K415 Issue No 1 C Closed O Open Figure B2 Test Connections Low Pressure Calibrator Versions ...
Страница 87: ...77 K415 Issue No 1 Appendix 1 DATA LOG FILE EXAMPLE ...
Страница 88: ...K415 Issue No 1 78 This page intentionally blank ...