Cal mode (DPI 615 versions only)
Cal mode, which is available in tasks P-I, P-P, P-V, P-P, P-DISPLAY and P-SWITCH, provides
a method of setting-up test parameters manually. Downloaded test procedures can also
automatically set-up and turn on the Cal Mode function. The method of turning on and
setting-up Cal Mode is shown below for a P-I task. The method is similar for all the other
tasks available in Cal Mode.
Pressing the F1 key (TURN ON CAL MODE), provides the set-up screen for the CAL mode.
Initially, the cursor is placed in the UUT SPAN field to allow the required span range to be
entered. The corresponding values for the UUT output parameter (current) are then set,
followed by the maximum error value and error type (% rdg or % span). When all test
parameters have been set-up, the screen changes to display the input and output and
the test results. The test result can only be displayed to within a range of ±9.99%. If the
test result is outside this range, either the left pointing (-ve error) or right pointing (+ve
error) chevrons are displayed. Within this error band, the actual tolerance value is
displayed. Test results can either be stored as snapshots or logged as datalog files,
depending upon how the instrument has been set-up.
Basic Mode (Task BASIC)
This instrument powers up in this mode the first time that it is used. To select
BASIC
from
any other task, press the
TASK
key and select
BASIC
and press the
ENTER
key.
BASIC
mode is fully described in the
Getting Started,
section (see page 7).
FCT.: P-I
ENREG DONNEES
FCT.: P-I
ENREG DONNEES
24V OFF
24V OFF
SELECTIONNER
VALEUR
COURANT
PRESSION INT
RENTRER EN
MODE CALIB.
CHANGER
VALEUR
SORTIE
LINEAIRE
TYPE D'ERREUR
MAX. ERROR
% ECH.
% ECH. REFUSE
SORTIR DU
MODE CALIB.
Task Selection
Содержание Druck DPI 610
Страница 1: ...Druck DPI 610 615 Portable Pressure Calibrator Series User manual K415 GE Sensing ...
Страница 2: ......
Страница 10: ...K415 Issue No 1 viii ...
Страница 69: ...59 K415 Issue No 1 Hydraulic Calibrator Versions ...
Страница 70: ...K415 Issue No 1 60 This page intentionally left blank ...
Страница 76: ...K415 Issue No 1 66 This page intentionally left blank ...
Страница 77: ...67 K415 Issue No 1 Low Pressure Calibrator Versions ...
Страница 78: ...K415 Issue No 1 68 This page intentionally left blank ...
Страница 81: ...71 K415 Issue No 1 C Closed O Open Figure B2 Test Connections Low Pressure Calibrator Versions ...
Страница 87: ...77 K415 Issue No 1 Appendix 1 DATA LOG FILE EXAMPLE ...
Страница 88: ...K415 Issue No 1 78 This page intentionally blank ...