45
K415 Issue No. 1
Using the Calibration Menu
The calibration routines are selected from the Settings menu as detailed on page 38.
Enter the calibration PIN number, initially set to
4321
, press
and the display shows
the Calibration Menu.
PIN security
To prevent unauthorised access, it is recommended that the PIN number be changed as
soon as possible.
Change PIN
To change the PIN code, select
CHANGE PIN
from the calibration menu and proceed as
follows:
Note:
To set and verify a new PIN, the new PIN code must be entered twice.
• If the second code entered differs from the first, the new PIN will not be set.
Calibrate Internal Ranges
Select the
INT RANGES
from the menu as shown above and follow the calibration
procedure on the display.
SELECT OPTION
CALIBRATION
EXT SENSORS
CHANGE PIN
INT RANGES
V v t
SET NEW PIN
SET NEW PIN AGAIN
Enter New PIN
5 4 3 2
Enter New PIN
ATTENTION
CODE NON VALIDE
SELECTIONNE
OPTION CAL
CALIBRATION
CAPTEURS EXT
MODIFIE CODE
GAMMES INTER
Calibration
Содержание Druck DPI 610
Страница 1: ...Druck DPI 610 615 Portable Pressure Calibrator Series User manual K415 GE Sensing ...
Страница 2: ......
Страница 10: ...K415 Issue No 1 viii ...
Страница 69: ...59 K415 Issue No 1 Hydraulic Calibrator Versions ...
Страница 70: ...K415 Issue No 1 60 This page intentionally left blank ...
Страница 76: ...K415 Issue No 1 66 This page intentionally left blank ...
Страница 77: ...67 K415 Issue No 1 Low Pressure Calibrator Versions ...
Страница 78: ...K415 Issue No 1 68 This page intentionally left blank ...
Страница 81: ...71 K415 Issue No 1 C Closed O Open Figure B2 Test Connections Low Pressure Calibrator Versions ...
Страница 87: ...77 K415 Issue No 1 Appendix 1 DATA LOG FILE EXAMPLE ...
Страница 88: ...K415 Issue No 1 78 This page intentionally blank ...