K415 Issue No. 1
56
Ambient Temperature Channel
Use the following procedure for calibrating the ambient temperature measurement
channel.
(1)
Switch the instrument on, enter the
CALIBRATION
mode and select
TEMPERATURE
:
(2)
Allow the instrument’s temperature to stabilize in the calibration environment for
at least one hour.
(3)
Read the environmental temperature on a calibrated digital thermometer and
enter the recorded value as shown below. Example shown for a measured
ambient temperature of 21.5° Celsius.
Note:
Only one temperature point is required.
(4)
Press the
ENTER
key to accept the calibration. Press the
EXIT
key once to return
to the calibration menu or four times to quit the
CALIBRATION
and
SETUP
modes.
CHANGER
VALEUR
DEMARRER
Entrer val appliquee:
CALIBRE TEMP AIR
-10 à 40°C
CHANGER
VALEUR
Entrer val appliquee:
+
CHANGER VALEUR
Etalonnage Accepte ?
OUI
NON
CALIBRE TEMP AIR
-10 à 40°C
CALIBRE TEMP AIR
-10 à 40°C
m = 2.49000e-04
c = -1.02
Applique
Mesure
21.5
623298
CALIBRE TEMP AIR
-10 à 40°C
RESULTATS
Calibration Terminee
App ENTER pour continuer
Calibration
Содержание Druck DPI 610
Страница 1: ...Druck DPI 610 615 Portable Pressure Calibrator Series User manual K415 GE Sensing ...
Страница 2: ......
Страница 10: ...K415 Issue No 1 viii ...
Страница 69: ...59 K415 Issue No 1 Hydraulic Calibrator Versions ...
Страница 70: ...K415 Issue No 1 60 This page intentionally left blank ...
Страница 76: ...K415 Issue No 1 66 This page intentionally left blank ...
Страница 77: ...67 K415 Issue No 1 Low Pressure Calibrator Versions ...
Страница 78: ...K415 Issue No 1 68 This page intentionally left blank ...
Страница 81: ...71 K415 Issue No 1 C Closed O Open Figure B2 Test Connections Low Pressure Calibrator Versions ...
Страница 87: ...77 K415 Issue No 1 Appendix 1 DATA LOG FILE EXAMPLE ...
Страница 88: ...K415 Issue No 1 78 This page intentionally blank ...