K415 Issue No. 1
58
(6)
Apply the full-scale pressure to the external transducer and store the full-scale
(FS) point.
(7)
Release the applied pressure and disconnect the pressure reference. Press the
ENTER
key to accept the calibration. Press the
EXIT
key once to return to the
calibration menu or four times to quit the
CALIBRATION
and
SETUP
modes.
(8)
Check the calibration of the external transducer by applying the test pressures as
detailed in Table 1, page 44.
Add External Sensor
To add an external sensor to the directory of external sensors, proceed as follows:
• Connect the required external transducer to the EXT TRANSDUCER socket
located on the rear of the instrument.
Digitally compensated transducers will be downloaded into the directory as soon as the
F1 key is operated. Data for other types can be edited by selecting
EDIT
(F2). If the data
cannot be edited a warning message (INVALID ACTION) is displayed.
CALIBRATING PRESSURE EXT
RANGE 20000
kPa g
CALIBRATING PRESSURE EXT
RANGE 20000
kPa g
CALIBRATING PRESSURE EXT
RANGE 20000
kPa g
Apply nominal FS value
Press ENTER key when
applied value is stable
+
CHANGE VALUE
Содержание Druck DPI 610
Страница 1: ...Druck DPI 610 615 Portable Pressure Calibrator Series User manual K415 GE Sensing ...
Страница 2: ......
Страница 10: ...K415 Issue No 1 viii ...
Страница 69: ...59 K415 Issue No 1 Hydraulic Calibrator Versions ...
Страница 70: ...K415 Issue No 1 60 This page intentionally left blank ...
Страница 76: ...K415 Issue No 1 66 This page intentionally left blank ...
Страница 77: ...67 K415 Issue No 1 Low Pressure Calibrator Versions ...
Страница 78: ...K415 Issue No 1 68 This page intentionally left blank ...
Страница 81: ...71 K415 Issue No 1 C Closed O Open Figure B2 Test Connections Low Pressure Calibrator Versions ...
Страница 87: ...77 K415 Issue No 1 Appendix 1 DATA LOG FILE EXAMPLE ...
Страница 88: ...K415 Issue No 1 78 This page intentionally blank ...