23
K415 Issue No. 1
Max/Min Process Function
To set-up an input display to show max/min and present input reading, enable
MAX/MIN from the process menu and press F1 (SETTINGS) to provide
RESET
function. The display now shows the max/min values as follows:
Reset Max/Min display at any time by pressing the
F1 key.
To quit max/min, press
INPUT
, select
MAX/MIN
from process menu and switch the function off.
Filter Process Function
To apply the low pass filter to a selected input, enable
FILTER
from the process
menu and press F1 (SETTINGS) to provide access to the filter parameters. Two
settings are required,
Time to Settle
and
Band
.
To examine the current filter settings and exit without change, press the
EXIT
key.
The set-up procedure is as follows:
FCT. : CONFIG
MODE ENREG. RAPIDE
24V OFF
ACTUEL
RESET
MAX/MIN
PERIODE
HEURE =
2
Seconds
BANDE = AMPLITUDE MAX
QUI DOIT FILTREE VARIATIONS
SUPERIEURESA LA BANDE DU
FILTRE
BANDE = 1.5
% FS
PERIODE
HEURE = 2
Seconds
BANDE =
1.5
% FS
CONFIG.FILTRE
CONFIG.FILTRE
SELECTIONNER
VALEUR
SELECTIONNER
VALEUR
BANDE = AMPLITUDE MAX
QUI DOIT FILTREE VARIATIONS
SUPERIEURESA LA BANDE DU
FILTRE
CHANGER
VALEUR
CHANGER
VALEUR
Advanced Task
Process Functions
Содержание Druck DPI 610
Страница 1: ...Druck DPI 610 615 Portable Pressure Calibrator Series User manual K415 GE Sensing ...
Страница 2: ......
Страница 10: ...K415 Issue No 1 viii ...
Страница 69: ...59 K415 Issue No 1 Hydraulic Calibrator Versions ...
Страница 70: ...K415 Issue No 1 60 This page intentionally left blank ...
Страница 76: ...K415 Issue No 1 66 This page intentionally left blank ...
Страница 77: ...67 K415 Issue No 1 Low Pressure Calibrator Versions ...
Страница 78: ...K415 Issue No 1 68 This page intentionally left blank ...
Страница 81: ...71 K415 Issue No 1 C Closed O Open Figure B2 Test Connections Low Pressure Calibrator Versions ...
Страница 87: ...77 K415 Issue No 1 Appendix 1 DATA LOG FILE EXAMPLE ...
Страница 88: ...K415 Issue No 1 78 This page intentionally blank ...