21
K415 Issue No. 1
Process Functions
If required, the following process functions are available on the input display but
only
in
ADVANCED
task. If the instrument is in any other mode i.e. BASIC or any other task
mode, the input and output displays must first be configured in
ADVANCED
task.
Note:
PROCESS functions are not available to the output channel.
A summary of the process functions follows:
Tare
Allows either the current display value or a manually
entered value to be tared off display parameter reading.
Max/Min
Displays running Max/Min and present display values
simultaneously. Resettable via F1 key.
Filter
Applies low pass filter function to displayed parameter.
Filter characteristics
(Settling time and Band)
are user
programmable.
Flow
Applies square root function to displayed parameter.
% Span
Converts displayed parameter reading to a percentage of
span. Span definable via the F1 key.
Following selection of
ADVANCED
from the task menu, press the
INPUT
key. Use the
up
or
down
cursor keys to select the required input. Press the
PROCESS
(F1) key and
use the
left
or
right
cursor keys to enable the process on/off:
Press
ENTER
to switch the process ON with existing settings or F1 to change process
settings (where applicable).
INPUT
SELECTIONNER
TRAITEMENT
VALIDER
TRAITEMENT
ENTR. PROCESS
TARE
ON
MAX/MIN
OFF
FILTRE
OFF
DEBIT
OFF
% ECH
OFF
SELECT. ENTREE
PRESSION
INT
TENSION
COURANT
PRESSION
EXT
TEMPERATURE
PROCESS
UNITES
PARAMETRAGES
Advanced Task
Process Functions
Содержание Druck DPI 610
Страница 1: ...Druck DPI 610 615 Portable Pressure Calibrator Series User manual K415 GE Sensing ...
Страница 2: ......
Страница 10: ...K415 Issue No 1 viii ...
Страница 69: ...59 K415 Issue No 1 Hydraulic Calibrator Versions ...
Страница 70: ...K415 Issue No 1 60 This page intentionally left blank ...
Страница 76: ...K415 Issue No 1 66 This page intentionally left blank ...
Страница 77: ...67 K415 Issue No 1 Low Pressure Calibrator Versions ...
Страница 78: ...K415 Issue No 1 68 This page intentionally left blank ...
Страница 81: ...71 K415 Issue No 1 C Closed O Open Figure B2 Test Connections Low Pressure Calibrator Versions ...
Страница 87: ...77 K415 Issue No 1 Appendix 1 DATA LOG FILE EXAMPLE ...
Страница 88: ...K415 Issue No 1 78 This page intentionally blank ...