17
K415 Issue No. 1
Leak Test (LEAK TEST) Task
A - External pressure source (indicator instruments only)
B - Pressure regulator
C - Unit/system under test
• If required, use the
INPUT
key to change pressure units.
• Set-up the leak test WAIT and DURATION times to the required values as shown
below. Recommended minimum wait period - 3 minutes.
• Close the vent valve and pressurize the unit/system to the required
LEAK
TEST
pressure.
• Press the
RUN (F2)
key to start the leak test. When completed, the beeper sounds
and the leak test results are written to the display.
bar
F2
F1
DEMARRER
CHANGER
VALEUR
ATTENTE
DUREE
DEMARRER
PRESS. FIN
FUITE
TAUX FUITE
min
min
Bar
bar
bar
bar/m
3
1
C
A
B
FCT. : TESTFUITE
24V OFF
MODE ENREG. RAPIDE
DPI 615
CION
ATTENTE
DUREE
DEMARRER
PRESS. FIN
FUITE
3
TAUX FUITE
min
min
bar
bar
bar
bar/m
1
PRESSION INT
bar
DEMARRER
FCT. : TESTFUITE
MODE ENREG. RAPIDE
CION
ATTENTE
DUREE
DEMARRER
PRESS. FIN
FUITE
3
TAUX FUITE
min
min
bar
bar
bar
bar/m
1
PRESSION INT
bar
FCT. : TESTFUITE
MODE ENREG. RAPIDE
24V OFF
24V OFF
CHANGER
VALEUR
CHANGER
VALEUR
DEMARRER
REGLER
DUREE
REGLER
PERIODE
ATTENTE
Taking Measurements
Содержание Druck DPI 610
Страница 1: ...Druck DPI 610 615 Portable Pressure Calibrator Series User manual K415 GE Sensing ...
Страница 2: ......
Страница 10: ...K415 Issue No 1 viii ...
Страница 69: ...59 K415 Issue No 1 Hydraulic Calibrator Versions ...
Страница 70: ...K415 Issue No 1 60 This page intentionally left blank ...
Страница 76: ...K415 Issue No 1 66 This page intentionally left blank ...
Страница 77: ...67 K415 Issue No 1 Low Pressure Calibrator Versions ...
Страница 78: ...K415 Issue No 1 68 This page intentionally left blank ...
Страница 81: ...71 K415 Issue No 1 C Closed O Open Figure B2 Test Connections Low Pressure Calibrator Versions ...
Страница 87: ...77 K415 Issue No 1 Appendix 1 DATA LOG FILE EXAMPLE ...
Страница 88: ...K415 Issue No 1 78 This page intentionally blank ...