K415 Issue No. 1
54
(4)
Measure the 10% full-scale output current and enter the value measured on the
external milliammeter:
(5)
Measure the 90% full-scale output current and enter the value measured on the
external milliammeter
(6)
Press the
ENTER
key to accept the calibration. Press the
EXIT
key four times to
quit the
CALIBRATION
and
SETUP
modes.
(7)
Verify the instrument’s output current calibration by setting up the loop test
currents as shown in Table 4 and checking the set values against the
milliammeter standard.
(8)
Press the
OUTPUT
key, select
24 VOLT
and turn the 24V supply to
ON
. Proceed as
follows:
Measure 90% FS - output
Press ENTER key when
applied value is stable
Store Cal point?
YES
NO
Enter Measured Current:
RUN
:
Calibration
Содержание Druck DPI 610
Страница 1: ...Druck DPI 610 615 Portable Pressure Calibrator Series User manual K415 GE Sensing ...
Страница 2: ......
Страница 10: ...K415 Issue No 1 viii ...
Страница 69: ...59 K415 Issue No 1 Hydraulic Calibrator Versions ...
Страница 70: ...K415 Issue No 1 60 This page intentionally left blank ...
Страница 76: ...K415 Issue No 1 66 This page intentionally left blank ...
Страница 77: ...67 K415 Issue No 1 Low Pressure Calibrator Versions ...
Страница 78: ...K415 Issue No 1 68 This page intentionally left blank ...
Страница 81: ...71 K415 Issue No 1 C Closed O Open Figure B2 Test Connections Low Pressure Calibrator Versions ...
Страница 87: ...77 K415 Issue No 1 Appendix 1 DATA LOG FILE EXAMPLE ...
Страница 88: ...K415 Issue No 1 78 This page intentionally blank ...