![GE Druck DPI 610 Скачать руководство пользователя страница 5](http://html.mh-extra.com/html/ge/druck-dpi-610/druck-dpi-610_user-manual_132657005.webp)
iii
K415 Issue No. 1
Specification
Safe working pressure
20 bar range (300 psi)
1.75 x full-scale
350 bar range (5000 psi)
1.2 x full-scale
400 bar range (6000 psi)
1.5 x full-scale
All other ranges
2 x full-scale
Accuracy
Combined non-linearity, hysteresis and repeatability
±70 mbar range (2 inHg)
0.05% F.S.
up to ±150 mbar (4.4 inHg)
0.05% span
200 mbar to 20 bar (3 psi to 300 psi) [Calibrator]):
0.025% F.S.
35 bar to 700 bar (500 psi to 10000 psi) [Indicator]
0.025% F.S.
70 bar to 400 bar (1000 psi to 6000 psi) [Hydraulic]
0.025% F.S.
Pressure Ranges
Refer to the pressure range matrix in the data sheet.
Temperature Effects
±0.004% of reading/°C (averaged over -10° to +40°C w.r.t. 20°C)
±0.002% of reading/°F (averaged over +14° to 104°F w.r.t. 68°F)
Power supply
Batteries
6 x 1.5 V C cells, alkaline (up to 60 hours nominal use at 20°C)
Rechargeable NiCad battery pack (20 hours nominal use) supplied with charger/
adaptor, supplies power to instrument while charging batteries.
Voltage Inputs
Range:
±50V
Accuracy
±0.05% rdg, ±0.004% F.S.
Resolution
100µV max
Voltage Outputs
Range:
±10V
Accuracy
±0.1%
Load
10mA
Range:
±24V
Accuracy
±5%
Load
26mA
Current Inputs
Range:
±55mA
Accuracy
±0.05% rdg, ±0.004% F.S.
Resolution
1µA max
Current Output
Range:
24mA
Accuracy
±0.05% rdg, ±0.01% F.S.
Resolution
1µA max
Содержание Druck DPI 610
Страница 1: ...Druck DPI 610 615 Portable Pressure Calibrator Series User manual K415 GE Sensing ...
Страница 2: ......
Страница 10: ...K415 Issue No 1 viii ...
Страница 69: ...59 K415 Issue No 1 Hydraulic Calibrator Versions ...
Страница 70: ...K415 Issue No 1 60 This page intentionally left blank ...
Страница 76: ...K415 Issue No 1 66 This page intentionally left blank ...
Страница 77: ...67 K415 Issue No 1 Low Pressure Calibrator Versions ...
Страница 78: ...K415 Issue No 1 68 This page intentionally left blank ...
Страница 81: ...71 K415 Issue No 1 C Closed O Open Figure B2 Test Connections Low Pressure Calibrator Versions ...
Страница 87: ...77 K415 Issue No 1 Appendix 1 DATA LOG FILE EXAMPLE ...
Страница 88: ...K415 Issue No 1 78 This page intentionally blank ...