1
K415 Issue No. 1
General
The versions of the DPI 610 and DPI 615 instruments are: pneumatic indicator, pneumatic
calibrator, hydraulic calibrator and low pressure pneumatic calibrator. All instruments
measure and display pneumatic and hydraulic pressure applied to the test port or to an
externally connected pressure sensor. Pressure measurement can be absolute, gauge
and sealed gauge and in ranges from 2.5 mbar to 700 bar (1.0 inH
2
O to 10000 psi).
The calibrator versions of this instrument contain pneumatic or hydraulic pressure
generation components to produce pneumatic pressure ranges between -1 to 20 bar
(-14.5 psi to 300 psi) and hydraulic pressure ranges up to 400 bar (6000 psi).
The electrical connections, on the front of the instrument, enable the instrument to
measure ±50 volts d.c. and ±55 mA and produce 10 volts d.c. or 24 volts d.c. and a
maximum of 24 mA. An integral sensor provides measurement of ambient temperature.
Additional sensors (option B1) connect to an external connector and extend the pressure
measurement range and include differential pressure measurement. The instrument has
an RS232 connector to enable uploading of test data to a compatible documenting
system. The DPI 615 has the ability to download, from a PC, pre-defined calibration and
test routines. Six alkaline C size batteries or (option A) rechargeable batteries with a
charger/adaptor, power the instrument.
Important Notice
Zinc-carbon and zinc-chloride cells should
NOT
be used in this instrument.
Use only the battery types as shown in the table on page 7.
Description of Procedures
In the procedures outlined in this user guide, hard (fixed function) and soft (variable
function) key operations are shown in bold type (e.g.)
TASK
and
F1
. These statements
mean press the
TASK
key and press the
F1
key. Soft key operations can be allocated to
both the F1 and F2 keys. Where a specific soft function is referred to it is written in bold
italics (e.g.)
PROCESS
.
This instrument has a number of operating modes that are described in a simplified form
in the following sections. Diagrams accompanying the procedures give typical selection
sequences and shaded controls indicate that this control key should be pressed in the
appropriate sequence. Diagrams should be read from left to right, top to bottom where
appropriate. A shaded display soft box indicates that the function key immediately
below that soft box should be pressed (either
F1
for the left hand soft box or
F2
for the
right).
In the above diagram the following key sequence is indicated.
(a)
Press the F2 key (the key immediately below the
PRESSURE UNITS
soft box).
(b)
Use the
Up
and
Down
cursor keys (only) to select the required option. (If all keys
shaded, use all these keys to select or enter data).
(c)
Press the
ENTER
key.
SELECTIONNER
UNITES
inHg
bar
hPa
Pa
bar
V
COURANT
UNITES
FCT.: STANDARD
+
-
CHOIX UNITES
DE PRESSION
INTRODUCTION
Summary of Functions
Содержание Druck DPI 610
Страница 1: ...Druck DPI 610 615 Portable Pressure Calibrator Series User manual K415 GE Sensing ...
Страница 2: ......
Страница 10: ...K415 Issue No 1 viii ...
Страница 69: ...59 K415 Issue No 1 Hydraulic Calibrator Versions ...
Страница 70: ...K415 Issue No 1 60 This page intentionally left blank ...
Страница 76: ...K415 Issue No 1 66 This page intentionally left blank ...
Страница 77: ...67 K415 Issue No 1 Low Pressure Calibrator Versions ...
Страница 78: ...K415 Issue No 1 68 This page intentionally left blank ...
Страница 81: ...71 K415 Issue No 1 C Closed O Open Figure B2 Test Connections Low Pressure Calibrator Versions ...
Страница 87: ...77 K415 Issue No 1 Appendix 1 DATA LOG FILE EXAMPLE ...
Страница 88: ...K415 Issue No 1 78 This page intentionally blank ...