57
K415 Issue No. 1
Calibrate External Sensors
Use the following procedure for calibrating external pressure sensors.
(1)
Connect the required external transducer to the EXT TRANSDUCER socket located
on the rear of the instrument.
(2)
Allow the instrument’s temperature and the temperature of the external
transducer to stabilize in the calibration environment for a minimum of 1 hour.
(3)
Switch the instrument on, enter
CALIBRATION
mode and select
EXT SENSORS
from the
CALIBRATION
menu:
(4)
Select the transducer to be calibrated from the transducer menu using the
cursor keys and press ENTER:
If the sensor to be calibrated is not in the directory, or no sensors are listed in the
directory, press the
F1
key (ADD NEW SENSOR). This will place the sensor in the
directory, allowing it to then be selected.
(5)
Connect the pressure standard to the inlet of the external transducer, apply the
zero point pressure and store the zero point:
INT RANGES
CHANGE PIN
EXT SENSORS
N/S
GAMME
bar
TYPE
N/S
GAMME
bar
TYPE
1154737
20.000
9
EDITION
N/S
GAMME
bar
TYPE
1154737
20.000
9
ENLEVER
CAPTEURS EXT
CAPTEURS EXT
CAPTEURS EXT
AJOUTER
NOUVEAU CAPT.
AJOUTER
NOUVEAU CAPT.
MODIFIER
DETAILS
Calibration
Содержание Druck DPI 610
Страница 1: ...Druck DPI 610 615 Portable Pressure Calibrator Series User manual K415 GE Sensing ...
Страница 2: ......
Страница 10: ...K415 Issue No 1 viii ...
Страница 69: ...59 K415 Issue No 1 Hydraulic Calibrator Versions ...
Страница 70: ...K415 Issue No 1 60 This page intentionally left blank ...
Страница 76: ...K415 Issue No 1 66 This page intentionally left blank ...
Страница 77: ...67 K415 Issue No 1 Low Pressure Calibrator Versions ...
Страница 78: ...K415 Issue No 1 68 This page intentionally left blank ...
Страница 81: ...71 K415 Issue No 1 C Closed O Open Figure B2 Test Connections Low Pressure Calibrator Versions ...
Страница 87: ...77 K415 Issue No 1 Appendix 1 DATA LOG FILE EXAMPLE ...
Страница 88: ...K415 Issue No 1 78 This page intentionally blank ...