19
K415 Issue No. 1
Relief Valve Test (REL VALVE) Task
To carry out a relief valve test, press
TASK
and select
REL VALVE
. Connect the output
pressure port of the instrument to an external system as shown below:
A - External pressure source (indicator instruments only)
B - Pressure regulator
C - Relief valve under test
•
To change the pressure units, if required, press
INPUT
and select the required
units using the cursor keys.
•
If necessary, turn on the 24 Volt internal supply by pressing
OUTPUT
, select
24
VOLT
and switch
ON
with the right cursor button and press
ENTER
.
•
Close the vent valve and, using the hand-pump or external pressure supply, apply
pressure to the relief valve under test.
•
When the relief valve operates, the maximum recorded pressure indicates the
operating point of the valve.
Note:
The
STORE
key can be used for this purpose. Use right cursor key initially,
followed by up/down keys to enter Snapshot text).
•
Record the test results.
•
Open vent valve to release test pressure.
Note:
If using external pressure supply, isolate supply before opening the vent
valve.
bar
bar
RESET
MAXIMUM
MINIMUM
MAX/MIN
F2
F1
FCT. : SOUPAPE
24V OFF
MODE ENREG. RAPIDE
DPI 615
C
A
B
Taking Measurements
Содержание Druck DPI 610
Страница 1: ...Druck DPI 610 615 Portable Pressure Calibrator Series User manual K415 GE Sensing ...
Страница 2: ......
Страница 10: ...K415 Issue No 1 viii ...
Страница 69: ...59 K415 Issue No 1 Hydraulic Calibrator Versions ...
Страница 70: ...K415 Issue No 1 60 This page intentionally left blank ...
Страница 76: ...K415 Issue No 1 66 This page intentionally left blank ...
Страница 77: ...67 K415 Issue No 1 Low Pressure Calibrator Versions ...
Страница 78: ...K415 Issue No 1 68 This page intentionally left blank ...
Страница 81: ...71 K415 Issue No 1 C Closed O Open Figure B2 Test Connections Low Pressure Calibrator Versions ...
Страница 87: ...77 K415 Issue No 1 Appendix 1 DATA LOG FILE EXAMPLE ...
Страница 88: ...K415 Issue No 1 78 This page intentionally blank ...