Troubleshooting
6-17
Exception
ID
Error
Warning
Exception
Description
(In system Log)
Message
(Displayed in User Interface)
Diagnosis Reference
0205503
Error
Attach shared
memory
segment failed
Memory exception. Please
restart the system. If this issue
persists, please call for service.
Check if there exists some
unwanted share memory in
the system, or check if
software is installed
correctly.
0205504
Error
Initialize and
open a named
semaphore
failed
Semaphore exception. Please
restart the system. If this issue
persists, please call for service.
Check if there exists some
unwanted semaphore in
the system, or check if
software is installed
correctly.
0205505
Error
Post a
semaphore
failed
Semaphore exception. Please
restart the system. If this issue
persists, please call for service.
Check if the semaphore is
valid, or check if the
maximum allowable value
for the semaphore will be
exceeded.
0205506
Warning
FIFO is full
Image FIFO full. The mainframe
stopped. Please restart the
system. If this issue persists,
please call for service.
Check CPU Affinity and
thread priority. Complied
SW version (Debug or
Release)
0205507
Warning
Load IP
configuration
file failed
IP configuration exception.
Image quality may be affected.
Please restart the system. If
this issue persists, please call
for service.
Check if IP configuration file
crashed
0205508
Warning
Load setup
configuration
file failed
Setup configuration exception.
Some system setting may be
missed. Please restart the
system. If this issue persists,
please call for service.
Check if Setup configuration
file crashed
0205509
Error
Init critical
section failed
Execution exception. Please
restart the system. If this issue
persists, please call for service.
System unknown error
0205510
Error
Lock critical
section failed
Execution exception. Please
restart the system. If this issue
persists, please call for service.
System unknown error
0205511
Error
Unlock critical
section failed
Execution exception. Please
restart the system. If this issue
persists, please call for service.
System unknown error
0205512
Warning
Init Xkey lock
failed
Image processing LEDs on
keyboard display exception.
The display may be abnormal.
Please ensure no key is
pressed on keyboard, then
press “Auto B/C” key and
“Negate” key. If this issue
persists, please call for service.
Check system IO state and
keyboard connection
Содержание Brivo OEC 715
Страница 2: ......
Страница 19: ...Chapter1 Introduction and Safety...
Страница 41: ...Introduction and Safety 23 46 54 20 18 26...
Страница 55: ...Chapter2 System Overview...
Страница 137: ...Chapter3 Installation...
Страница 147: ...Installation 3 11 7 Put the ramp in position Release all bracket and move the workstation along the ramp off the pallet...
Страница 211: ...Installation 3 75 3 9 Reporting For U S Systems Complete and submit Form 2579 within 15 days of installation...
Страница 212: ...Chapter4 Calibration...
Страница 223: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 12 Mainframe Control Board TP7...
Страница 231: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 20 Mainframe Control Board TP17 POT4 J5 TP7 POT7 POT5...
Страница 275: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 64...
Страница 284: ...Chapter5 Software...
Страница 312: ...Software 5 29 4 Click Next 5 Click Install 6 Click Finish button to complete installation procedure...
Страница 326: ...Software 5 43 2 Click on install to continue 3 Click Next to continue...
Страница 335: ...Chapter6 Troubleshooting...
Страница 408: ...Chapter7 Replacement...
Страница 418: ...Replacement 7 11 166...
Страница 488: ...Chapter8 Periodic Maintenance...
Страница 502: ...Periodic Maintenance 8 15...
Страница 505: ...Chapter9 Technical Reference...
Страница 520: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 9 16 Lateral configuration 1 5m Lateral configuration 1m...
Страница 521: ...Technical Reference 9 17 Vertical configuration 1 5m Vertical configuration 1m...
Страница 526: ...11 Appendix System Schematics...